62 | Drystar 5500, Drystar 5503 | Advanced Operation (key-operator mode)
Density response of film
Different densitometers can give different results when measuring the same
density area. This is due to spectral differences between the densitometers.
The density values used in this manual are based on the X_Rite 310
densitometer. Values measured with another densitometer can vary from
these values.
It's possible to change the default densitometer. You can also compare the
density values from different densitometers with each other.
Non-calibrated or badly calibrated densitometers can result in even larger
differences.
Related Links
Film density response data
Changing general image quality settings - Default densitometer
2900J EN 20180320 1448
WARNING:
The calibration film must be free of dust particles or any
other artifacts. If this is not the case, clean the print head
and restart the calibration procedure.
on page 145
on page 41