Drystar 5500, Drystar 5503 | Advanced Operation (key-operator mode) | 61
• Calibration successful, but Dmax differs more than 2% from the target
value.
The obtained Dmax (x.xx) after calibration is displayed:
• Calibration procedure succesful, but limited by film-Dmax. The
obtained Dmax does not meet the required system Dmax and is limited
by the film sensitometry.
The obtained Dmax (x.xx) after calibration is displayed:
5. Press the Confirm key to return to the Calibration menu.
WARNING:
When the calibration process is interrupted, you will have to
start the calibration procedure again. Otherwise, the old
calibration parameters will be used.
6. Check the printed test film.
The test film must be similar to the image below.
2900J EN 20180320 1448