Honeywell Vertex Edge User Manual page 86

72-point continuous monitor
Hide thumbs Also See for Vertex Edge:
Table of Contents

Advertisement

Adjust the Flow Rate
With the system vacuum level set, the unit is now ready to adjust flow for all the points.
1.
If still in the Vacuum level screen, press back.
If not, enter the Adjust flow screen as described in steps 1-4 in theAdjust Supply Vacuum
section.
2.
Check the flow rate, and if there are any points marked in yellow, the point must be
adjusted. See below as an example.
3.
Tap RELEASE to access to the needle valve on the side of the Analyzer if it is locked to
access the needle valves.
4.
Turn the needle valve for the point that is to be adjusted until the flow reaches the target
flow rate of 200 ± 10cc/min. Once the flow is in this range, the bar will change color to
green.
5.
Repeat for all points in the analyzer that are out of range.
6.
Repeat for all analyzers until the entire system is complete.
7.
Once complete, press STOP FLOW for each analyzer. This will stop for the flow for each
analyzer.
8.
Once all analyzers have the flow stopped, the pump will turn off.