Honeywell Vertex Edge User Manual page 143

72-point continuous monitor
Hide thumbs Also See for Vertex Edge:
Table of Contents

Advertisement

2.
Locate the Ethernet cable on the rear panel and disconnect it.
3.
Open the side of the Analyzer by unscrewing the 4 thumbscrews at the top of the left side.
4.
Approximately 20 seconds after the Ethernet Cable has been disconnected, the Analyzer
will recognize that it has lost communications with the DAq and activate the "Manual
Override" button. The LED's on the front of the Analyzer will flash to show a Maintenance
Fault per the LED status flash pattern.
To reset faults and alarms press and hold button for 1-3 seconds.
To Exit Monitor and Open Gate press and hold button for 10 seconds and above
To restart monitoring mode, press and hold button for 4 to 9 seconds.
5.
Close and reattach the Analyzer cover, reconnect the Ethernet Communications Cable
and slide the Analyzer into the cabinet.
Fix an Unresponsive Vertex Edge Touch Screen
Completely resetting the computer can resolve many issues that cause a frozen/unresponsive
screen. Follow these steps to perform a hard reset:
1.
Remove any USB devices from the USB ports of the HMI PC if non-default USB devices
were inserted. The USB connection from touch screen should remain.
2.
Turn off the computer by pressing and holding the power button for a few seconds.
Vertex Edge System
142
User Manual