11. When new Chemcassette is installed correctly, tap NEXT.
12. The Analyzer reads the RFID tag on the Chemcassette tape and shows the tape information.
Check the Chemcassette information and Tap NEXT.
13. If the Chemcassette is not a brand-new tape, the user can accept the Chemcassette tape or
adjust Optics. If Accept button is pressed, it will accept the Chemcassette and update the
tape information without optics adjustment. If tape is not new out of the bag, a leader tape
can be used to run the Optics Adjust by pressing Adjust Optics. The leader tape will need to
come from a fully used tape. Do not rewind the CC to reuse the leader tape. If using the
leader tape, the user will need to manually install the leader tape for each color to be tested.
Vertex Edge System
101
User Manual