Optics drive unusually low
211
Optics drive unusually high
212
Excess optics signal
213
Double Pump failure
214
Flow gone for the specific point
Failed to stop Analyzers monitoring
215
mode
216
Point Pressure Out Of Range
217
Supply Pressure Out Of Range
218
Inadequate Analyzer Vacuum
219
Optics SW corrupted
220
Optics Internal SW errors
LED adjustment failed due to
221
insufficient optical signal
Vertex Edge System
Possible Cause
Optics LED not properly calibrated
Optics board defective
Analyzer CPU defective
Optics LED not properly calibrated
LED degradated
Optics board defective
Analyzer CPU defective
Optics LED not properly calibrated
Optics board defective
Analyzer CPU defective
See causes for Fault 109
Circuit breaker tripped
Improper flow setup
Tape tracking problem
Condensation in system
Clogged micro tube
Gate not fully closing
Optics block loose
Flow adjustment is unstable during adjusting
Queen Analyzer (first in the rack) failed to communicate with other
Analyzers
Analyzer hardware failure
Miscalibrated sensor board or defective transducer
Miscalibrated sensor board or defective transducer
Improper system pressure adjustment
Hardware failure
Cable issue
Hardware failure
Optic block dirty
Perform Replace CC or Adjust Optics operation to recalibrate
Replace the Optics Block. Contact Honeywell service
Replace Analyzer, Contact Honeywell service
Perform Replace CC or Adjust Optics operation to recalibrate
Replace the optics block
Replace the Optics Block. Contact Honeywell service
Replace Analyzer, Contact Honeywell service
Perform Replace CC or Adjust Optics operation to recalibrate
Replace the Optics Block. Contact Honeywell service
Replace Analyzer, Contact Honeywell service
See recoveries for Fault 109
Contact Honeywell service
Adjust flow to 200cc
Reload Chemcassette
Purge internal lines
Replace tube.
Note: Calibration required, contact Service.
Gate adjustment loose, pivot binding
Tighten optics block fasteners
Replace needle valves
Replace first Analyzer in the rack
Replace Analyzer, Contact Honeywell service
Replace Analyzer
Replace Analyzer
Adjust system pressure to 11 inHg
Replace the Optics Block
Confirm cable connections
Replace the Optics Block. Contact Honeywell service
Clean optics.
Contact Honeywell Analytics Service.