The Musica Acquisition Workstation Is Connected To The Dr Detector, But The Dr Detector Is Not Active (Access Point Mode) - AGFA XD 14 User Manual

Hide thumbs Also See for XD 14:
Table of Contents

Advertisement

118 | XD 14, XD⁺14 | Problem solving
The MUSICA Acquisition Workstation is connected to
the DR detector, but the DR detector is not active
(access point mode)
Details
Cause
Brief Sol-
ution
0430A EN 20210426 1355
The DR detector is configured for access point mode. The Wifi
settings in Windows show that the MUSICA Acquisition Worksta-
tion is connected to the DR detctor, but the DR Detector Switch
shows an error status for the DR detector.
Another MUSICA Acquisition Workstation is still connected to
the DR detector.
On the other MUSICA Acquisition Workstation, switch to the
wireless hospital network to disconnect from the DR detector.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Xd+14Fxrd-3643vawFxrd-3643vaw plus

Table of Contents