Step 4: check the exposure settings

On the DR Detector Switch:
1. Check if the DR Detector Switch displays the name of the DR Detector
that's being used
2. If a wrong DR Detector is displayed, select the right DR Detector by
clicking the drop down arrow on the DR Detector Switch.
3. Check the DR Detector Status icon.
On the X-ray system:
1. Check if the exposure settings displayed on the console are suitable for the
exposure.
2. Check if no error messages are displayed on the X-ray system.
Exposure synchronization
Depending on the configuration, the DR Detector synchronizes to the
exposure using one of these methods:
• X-ray generator synchronization
• Automatic exposure detection
Related Links
Automatic exposure detection
WARNING:
In a configuration using automatic exposure detection, the X-ray
system allows executing an exposure, even if the DR Detector is
not ready. Avoid unnecessary dose by checking the status of the
DR Detector before exposure. The DR Detector Switch displays
the DR Detector status icon.
on page 94
XD 14, XD⁺14 | Getting started | 81
0430A EN 20210426 1355