124 | XD 14, XD⁺14 | Technical Data
Atmospheric pressure
Environmental conditions (during storage and transport)
Temperature (ambient)
Humidity (non condensing)
Atmospheric pressure
Dimensions
Dimensions
width x length x height
Weight XD 14
Weight XD⁺14
Load
Maximum load
Maximum total load
Vibration tolerance
Shock tolerance
Drop limit
Image acquisition
Image acquisition time (mini-
mum cycle time)
Conversion screen
Pixel size
Active pixel matrix
Effective pixel matrix
Detector type
Active area size
Effective area size
0430A EN 20210426 1355
between 700 hPa and 1060 hPa
between -15 °C and +55 °C
between 5% and 90%
(non-condensing)
between 500 and 1060 hPa
approx. 384.0 x 460.0 x 15.0 mm
2.95 kg (incl. one battery)
3.15 kg (incl. two batteries)
3.10 kg (incl. one battery), 3.30 kg (incl. two
batteries)
200 kg on an area of 40 mm in diameter
400 kg over the whole detector surface
2 G during normal operation
5G during storage and transport
20 G during normal operation
30 G during storage and transport
1000 mm (once)
4 s
CsI
140 µm
2560 x 3072
2536 x 3048
amorphous silicium
258 mm x 430 mm
355.0 mm x 426.7 mm