76 | XD 14, XD⁺14 | Getting started

Starting the DR detector

To start the DR detector:
1. If the DR detector is connected to the System Control Unit via the DR
detector cable, check if the power cable of the System Control Unit is
connected to the mains power.
2. Turn on the System Control Unit using the power switch.
The status indicator is green.
Skip to step 6.
Depending on the configuration, the System Control Unit may not be part
of the system.
3. Fully charge the DR detector.
Charge the DR detector on the day of examination or on the previous day.
4. Take the DR detector out of the cradle.
Figure 15: Pull the DR detector forward to take it out of the cradle
5. Turn on the detector.
Press and hold the power button for 1 second.
0430A EN 20210426 1355
Note: The battery slowly discharges even if not in use. The
battery pack may have expired if it discharges immediately
after being fully charged. You can purchase a new battery pack
to replace an exhausted one.
WARNING:
When taking the DR detector out of the cradle by lifting it
upwards, shock may be applied to the detector. Always take
the DR detector out of the cradle by pulling it forward.
Note: Before operating the detector, start up the MUSICA
Acquisition Workstation.