| XD 17, XD⁺17 | Contents Contents Legal Notice ................5 Introduction to this Manual ........... 6 Scope .................7 About the safety notices in this document ....8 Disclaimer ..............9 Introduction to the DR Detector ........... 10 Intended use ............
Page 3
XD 17, XD⁺17 | Contents | iii Additional labeling of the DR detector dual battery charger ............47 Additional Labeling of the System Control Unit Additional Labeling of the Mini System Control Unit ............. 49 Additional Labeling of the System Control Unit Lite ..............
Page 4
| XD 17, XD⁺17 | Contents Connection status ........99 Charging a battery ..........100 Charging the DR detector in the DR detector cradle ............101 Charging the DR detector using the power adapter ................102 Charging the battery using the DR detector cable ................103...
Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium or its affiliates. XD 17 and XD⁺17 are trademarks of Agfa NV, Belgium or one of its affiliates. All other trademarks are held by their respective owners and are used in an editorial fashion with no intention of infringement.
6 | XD 17, XD⁺17 | Introduction to this Manual Introduction to this Manual Topics: • Scope • About the safety notices in this document • Disclaimer 0434A EN 20210426 1355...
XD 17, XD⁺17 | Introduction to this Manual | 7 Scope This manual contains information for the safe and effective operation of the XD 17 and XD⁺17 wireless DR detectors and peripheral equipment, further referred to as the DR detector. 0434A EN 20210426 1355...
8 | XD 17, XD⁺17 | Introduction to this Manual About the safety notices in this document The following samples show how warnings, cautions, instructions and notes appear in this document. The text explains their intended use. DANGER: A danger safety notice indicates a hazardous situation of direct, immediate danger for a potential serious injury to a user, engineer, patient or any other person.
XD 17, XD⁺17 | Introduction to this Manual | 9 Disclaimer Agfa assumes no liability for use of this document if any unauthorized changes to the content or format have been made. Every care has been taken to ensure the accuracy of the information in this document.
10 | XD 17, XD⁺17 | Introduction to the DR Detector Introduction to the DR Detector Topics: • Intended use • Indications for use of the DR Retrofit solution • Intended User • Configuration • Equipment Classification • Options and Accessories •...
Intended use Indications for use The XD 17 and XD⁺17 DR detectors are a digital X-ray imaging solution. They acquire images by detecting X-rays that has been passed through the human body. When X-ray photons pass through the scintillator in the detector, the photons convert to visible ray, and the visible ray is converted to electronic signals through TFTs –...
12 | XD 17, XD⁺17 | Introduction to the DR Detector Indications for use of the DR Retrofit solution The DR Retrofit solution is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of human anatomy. The DR Retrofit solution may be used wherever conventional screen-film systems may be used.
XD 17, XD⁺17 | Introduction to the DR Detector | 13 Intended User This manual is written for trained users of Agfa products. Users are considered as the persons who actually handle the equipment as well as the persons having authority over the equipment. Before attempting to work with this equipment, the user must read, understand, note and strictly observe all warnings, cautions and safety markings on the equipment.
14 | XD 17, XD⁺17 | Introduction to the DR Detector Configuration The DR detector is a component that can be integrated in an X-ray system and that communicates to a workstation. Multiple DR detectors can communicate to a single MUSICA Acquisition Workstation. A DR detector can be used on more than one MUSICA Acquisition Workstation.
Page 15
XD 17, XD⁺17 | Introduction to the DR Detector | 15 Related Links Managing network connections in client mode configuration on page 106 Managing network connections in access point mode configuration on page 108 0434A EN 20210426 1355...
16 | XD 17, XD⁺17 | Introduction to the DR Detector Equipment Classification Per EN/IEC60601-1, Medical Electrical Equipment, General Requirements for Safety, the DR Detector, including the battery pack, is classified as following. Type of protection Internally powered (wireless configuration)
Anti-scatter grids are used to reduce scattered radiation and improve image quality. Grids are available as an option. Refer to the Agfa website for specifications on the anti-scatter grids that have been found compatible with the system and the DR Detectors.
18 | XD 17, XD⁺17 | Introduction to the DR Detector Operation Controls Topics: ⁺ 17 • XD 17, XD • DR detector cradle • DR Detector dual battery charger • System Control Unit • System Control Unit Lite •...
Page 19
XD 17, XD⁺17 | Introduction to the DR Detector | 19 XD 17, XD⁺17 1 2 3 5 6 7 Power button with indicator light • Press to start the DR detector or to display the status of the DR detector.
Page 20
20 | XD 17, XD⁺17 | Introduction to the DR Detector Effective imaging area border and center position indication Figure 2: DR detector operation controls Knobs securing the battery cover plate. Cover plate for the DR detector battery The battery supplies power to the detector during wireless communication.
XD 17, XD⁺17 | Introduction to the DR Detector | 21 DR detector cradle The battery cradle has two slots to insert a DR detector. Slot A Slot B Power status indicator Status indicators for slot A and slot B •...
22 | XD 17, XD⁺17 | Introduction to the DR Detector DR Detector dual battery charger The dual battery charger is available depending on the configuration of the system. The battery charger has two slots to insert a battery. Connector for power adapter...
XD 17, XD⁺17 | Introduction to the DR Detector | 23 System Control Unit The System Control Unit is connected to the DR Detector via wireless network or via the DR Detector cable. The System Control Unit is connected to the X-ray generator to synchronize the exposure, in a configuration with X-ray generator synchronization.
24 | XD 17, XD⁺17 | Introduction to the DR Detector System Control Unit Lite The System Control Unit Lite is connected to the DR Detector via the DR Detector cable. The System Control Unit Lite is connected to the X-ray generator to synchronize the exposure, in a configuration with X-ray generator synchronization.
XD 17, XD⁺17 | Introduction to the DR Detector | 25 DR detector cable The DR detector cable connects the DR detector to the System Control Unit. The DR detector cable can be used for charging the battery of the DR detector, to power the DR detector and to transmit image data.
26 | XD 17, XD⁺17 | Introduction to the DR Detector DR Detector Switch The DR Detector Switch is available in the title bar of the MUSICA Acquisition Workstation. The DR Detector Switch shows which DR detector is active and shows its status. The DR Detector Switch can be used to activate another DR detector.
Page 27
XD 17, XD⁺17 | Introduction to the DR Detector | 27 Meaning DR detector DR detector is DR detector is DR detector is is ready for initializing for off or discon- inactive (no exposure exposure nected or in thumbnail se-...
28 | XD 17, XD⁺17 | Introduction to the DR Detector System Documentation The documentation consists of a User manual (this document) and related documentation: • MUSICA Acquisition Workstation user manual (document 4420). • MUSICA Acquisition Workstation key user manual (document 4421).
XD 17, XD⁺17 | Introduction to the DR Detector | 29 Training The user must have received adequate training on the safe and effective use of the system before attempting to work with it. Training requirements may vary from country to country. The user must make sure that training is received in accordance with local laws or regulations that have the force of law.
30 | XD 17, XD⁺17 | Introduction to the DR Detector Product Complaints Any health care professional (for example a customer or a user) who has any complaints or has experienced any dissatisfaction with the quality, durability, reliability, safety, effectiveness, or performance of this product must notify Agfa.
Compatibility The system must only be used in combination with other equipment or components if these are expressly recognized by Agfa as compatible. A list of such equipment and components is available from Agfa service on request. Changes or additions to the equipment must only be carried out by persons authorized to do so by Agfa.
32 | XD 17, XD⁺17 | Introduction to the DR Detector Compliance Topics: • General • Safety • Electromagnetic Compatibility • Radio Frequency 0434A EN 20210426 1355...
XD 17, XD⁺17 | Introduction to the DR Detector | 33 General • The product has been designed in accordance with the MEDDEV Guidelines relating to the application of Medical Devices and have been tested as part of the conformity assessment procedures required by 93/42/EEC Medical Device Directive (European Council Directive 93/42/EEC on Medical Devices).
Page 34
34 | XD 17, XD⁺17 | Introduction to the DR Detector • Local regulations • Restrictions on outdoor use • Specific absorption rate (SAR) Local regulations This product complies with local radio frequency regulations in the country or region where you purchased the product. Note that it cannot be used in any areas other than the country or region of its purchase.
36 | XD 17, XD⁺17 | Introduction to the DR Detector Wireless Communication Wireless communication is established between the internal wireless module of the DR detector and the MUSICA Acquisition Workstation via the wireless access point. The DR detector is compliant with IEEE 802.11n/ac (2.4 GHz/5 GHz).
XD 17, XD⁺17 | Introduction to the DR Detector | 37 Wired communication The use of accessories and cables other than those specified or sold by the manufacturer as replacement parts, may result in increased radiation emissions or decreased stability of the equipment.
38 | XD 17, XD⁺17 | Introduction to the DR Detector Installation Installation and configuration is performed by an Agfa trained and authorized service engineer. Contact your local support organization for more information. On a configuration with multiple DR detectors of the same type, it is required to apply labeling to the DR detector containing a unique nickname for each DR detector.
Page 39
XD 17, XD⁺17 | Introduction to the DR Detector | 39 Do not use this equipment in combination with peripherals such as defibrillators or large electric motors as these may cause power-supply noise or power supply voltage variations. Doing so may prevent normal operation of this equipment and peripherals.
40 | XD 17, XD⁺17 | Introduction to the DR Detector Messages Under certain conditions the DR detector shows a dialog box containing a message in the middle of the screen of the MUSICA Acquisition Workstation. This message informs the user that either a problem has occurred or that a requested action cannot be performed.
XD 17, XD⁺17 | Introduction to the DR Detector | 41 Labels Symbol Explanation On (power: connection to the mains) On (power: connection to the mains) for part of the equipment Off (power: disconnection from the mains) Off (power: disconnection from the mains) for part of...
Page 42
42 | XD 17, XD⁺17 | Introduction to the DR Detector Symbol Explanation Manufacturer Date of manufacture Serial number This mark shows compliance of the equipment with Di- rective 93/42/EEC (for European Union). This mark shows compliance with both Canadian and U.S.
Page 43
XD 17, XD⁺17 | Introduction to the DR Detector | 43 Symbol Explanation Dangerous voltage General Mandatory action. Topics: • Additional labeling of the DR detector • Additional labeling of the DR detector battery • Additional labeling of the DR detector cradle •...
44 | XD 17, XD⁺17 | Introduction to the DR Detector Additional labeling of the DR detector Type label on the back side of the DR detector. DR Detector identification label Label Meaning Writable label to identify and dedicate a DR Detector to an X- ray system bucky.
XD 17, XD⁺17 | Introduction to the DR Detector | 45 Additional labeling of the DR detector battery Type label on the back side of the battery. 0434A EN 20210426 1355...
46 | XD 17, XD⁺17 | Introduction to the DR Detector Additional labeling of the DR detector cradle Type label on the bottom side of the DR detector cradle. 0434A EN 20210426 1355...
XD 17, XD⁺17 | Introduction to the DR Detector | 47 Additional labeling of the DR detector dual battery charger Type label on the bottom side of the DR detector dual battery charger. 0434A EN 20210426 1355...
48 | XD 17, XD⁺17 | Introduction to the DR Detector Additional Labeling of the System Control Unit Type label on the bottom side of the System Control Unit. 0434A EN 20210426 1355...
XD 17, XD⁺17 | Introduction to the DR Detector | 49 Additional Labeling of the Mini System Control Unit Type label on the bottom side of the Mini Sys- tem Control Unit. 0434A EN 20210426 1355...
50 | XD 17, XD⁺17 | Introduction to the DR Detector Additional Labeling of the System Control Unit Lite Type label on the bottom side of the System Control Unit Lite. 0434A EN 20210426 1355...
XD 17, XD⁺17 | Introduction to the DR Detector | 51 Consulting the About box 1. Click About the solution in the Tools section of the Main Menu window on the MUSICA Acquisition Workstation. Figure 11: Main Menu window. This will open the About box showing the current release and version details of the DR Retrofit solution.
52 | XD 17, XD⁺17 | Introduction to the DR Detector Cleaning and Disinfecting All appropriate policies and procedures should be followed to avoid contamination of the staff, patients and equipment. All existing universal precautions should be extended to avoid potential contaminations and to avoid patients coming into (close) contact with the device.
XD 17, XD⁺17 | Introduction to the DR Detector | 53 Cleaning To clean the exterior of the equipment: 1. Stop the system WARNING: When the equipment is going to be cleaned, be sure to turn OFF the power of each device, and to unplug the power cord from the AC outlet.
54 | XD 17, XD⁺17 | Introduction to the DR Detector Use of protective plastic bag WARNING: Liquids ingressing the DR Detector may cause malfunction and contamination. If there is a chance that the detector comes in contact with liquids (bodily fluids, disinfectants,...), the DR Detector must be wrapped in a protective...
If you plan to use other disinfectants, approval of Agfa is needed before use, as most disinfectants can damage the device. UV disinfection is also not allowed. Perform the procedure following the instructions for use, the disposal instructions and the safety instructions of the selected disinfectants and tools and of the hospital.
56 | XD 17, XD⁺17 | Introduction to the DR Detector Approved disinfectants Refer to the Agfa website for specifications on the disinfectants that have been found compatible with the cover material of the device and can be used on the outer surface of the device.
XD 17, XD⁺17 | Introduction to the DR Detector | 57 Safety directions for disinfection WARNING: When the equipment is going to be cleaned, be sure to turn OFF the power of each device, and to unplug the power cord from the AC outlet.
58 | XD 17, XD⁺17 | Introduction to the DR Detector Maintenance Always consult the Agfa Service documentation and an Agfa trained and authorized service engineer for complete maintenance schedules. In order to ensure that the equipment is used safely and normally, be sure to inspect the equipment before use.
XD 17, XD⁺17 | Introduction to the DR Detector | 59 Daily inspection WARNING: For safety reasons, be sure to turn OFF the power to each piece of equipment before performing the following. Otherwise, an electric shock may result. 1. Ensure that cables are not damaged and cable jackets are not torn.
60 | XD 17, XD⁺17 | Introduction to the DR Detector Regular Inspection and Maintenance In order to ensure the safety of patients, operating personnel and third parties, and to maintain the performance and reliability of the equipment, be sure to perform regular inspection at least once a year. Clean up the equipment, make adjustments, or replace consumables.
XD 17, XD⁺17 | Introduction to the DR Detector | 61 Replacement Parts Support Parts required to maintain the functioning of the product will be stocked for seven years after discontinuance of production, to allow for repair. 0434A EN 20210426 1355...
(ISP), set in compliance with point 17(4) and 18(8) of Annex I of the EU Medical Device Regulation 2017/745, must be implemented and used in connection with the use of the Agfa medical device by the Customer (User). These are minimum requirements and designed to protect against unauthorised access that could hamper the device from functioning as intended.
Page 64
64 | XD 17, XD⁺17 | Introduction to the DR Detector • Medical devices shall be on an isolated network segment that restricts communication of the medical devices to the systems that are required for the device to function. • Internal firewalls shall be put in place to improve upon network segmentation and to further restrict communications of medical devices to the systems (internal and external) that they need to interact with.
XD 17, XD⁺17 | Introduction to the DR Detector | 65 Environmental Protection Disposal of this product in an unlawful manner may have a negative impact on health and on the environment. Therefore, when disposing of this product, be absolutely sure to follow the procedure which is in conformity with the laws and regulations applicable in your area.
66 | XD 17, XD⁺17 | Introduction to the DR Detector Safety Directions WARNING: Safety is only guaranteed when an Agfa certified field service engineer has installed the product. WARNING: Improper changes, additions, maintenance or repair of the system can lead to personal injury, electrical shock and damage to the equipment.
Page 67
XD 17, XD⁺17 | Introduction to the DR Detector | 67 may result in fire or electric shock. If liquid or water flows into an electrical component, turn off the power, mark it as "Out of Order" and contact service.
Page 68
68 | XD 17, XD⁺17 | Introduction to the DR Detector CAUTION: Handle the equipment carefully. Do not submerge the equipment in water. The internal image sensor may be damaged if something hits against it, or if it is dropped, or receives a strong jolt.
Page 69
XD 17, XD⁺17 | Introduction to the DR Detector | 69 2. Perform a calibration of the DR Detector. For instructions, refer to the DX-D DR Detector Calibration Key User Manual (document 0134). 3. Perform a flat field exposure and check the image for visible artifacts.
70 | XD 17, XD⁺17 | Introduction to the DR Detector Safety directions for the power supply WARNING: Do not operate the equipment using any type of power supply other than the one indicated on the rating label. Otherwise, it may result in fire or electric shock.
Page 71
XD 17, XD⁺17 | Introduction to the DR Detector | 71 WARNING: Do not connect the AC or DC power cable to the product with the power applied. Failure to do so may result in damage to the product. WARNING: Be sure to hold the plug or connector to unplug the power cord.
72 | XD 17, XD⁺17 | Introduction to the DR Detector Safety directions for the System Control Unit WARNING: Do not block the ventilation ports to prevent overheating. Overheating can cause system malfunction and damages. WARNING: Ensure continuous power supply to the system, with voltage and current according to the product specifications.
XD 17, XD⁺17 | Introduction to the DR Detector | 73 Safety directions for the DR detector battery WARNING: Do not use any means for charging the battery other than that specifically provided for use with the equipment. The battery is used with the DR Detector. Do not use them in other combinations.
Page 74
74 | XD 17, XD⁺17 | Introduction to the DR Detector Stop using the battery pack immediately if it emits smoke, a strange smell, or otherwise behaves abnormally. Do not let the battery pack and battery charger come into contact with water or other liquids and do not allow them to get wet.
76 | XD 17, XD⁺17 | Getting started Starting the DR detector To start the DR detector: 1. If the DR detector is connected to the System Control Unit via the DR detector cable, check if the power cable of the System Control Unit is connected to the mains power.
Page 77
XD 17, XD⁺17 | Getting started | 77 Figure 16: Power button During startup the power indicator is blinking green. After startup the power indicator is green. 6. Check the DR detector status icon on the DR Detector Switch. If the displayed status is error, following step is required: •...
XD 17, XD⁺17 | Getting started | 79 Step 1: retrieve the patient info At the MUSICA Acquisition Workstation: 1. When a new patient comes in, define the patient info for the exam. 2. Start the exam. If the workstation is connected to a second monitor that is positioned outside the operator room, make sure that the patient data is not exposed to unauthorized persons.
80 | XD 17, XD⁺17 | Getting started Step 3: prepare the exposure In the examination room: 1. Position the DR Detector. When using the bucky, check that the identification labels on the DR Detector and on the bucky match. Do not use a DR Detector that is dedicated to another bucky.
XD 17, XD⁺17 | Getting started | 81 Step 4: check the exposure settings On the DR Detector Switch: 1. Check if the DR Detector Switch displays the name of the DR Detector that's being used 2. If a wrong DR Detector is displayed, select the right DR Detector by clicking the drop down arrow on the DR Detector Switch.
82 | XD 17, XD⁺17 | Getting started Step 5: execute the exposure Press the exposure button to execute the exposure. Make sure the generator is ready for exposure before you press the exposure button. WARNING: The radiation indicator on the control console lights up during exposure release.
XD 17, XD⁺17 | Getting started | 83 Step 6: perform a quality control At the MUSICA Acquisition Workstation: 1. Select the image on which quality control is to be performed. 2. Prepare the image for diagnosis by using e.g. L/R markers or annotations.
84 | XD 17, XD⁺17 | Getting started Positioning the XD 17, XD⁺17 WARNING: Because the equipment cable is long, take care that cables do not become tangled during use. Also, be careful not to get your feet caught in the cable. It may cause a malfunction of the equipment or injury to the user from tripping over the cable.
Page 85
XD 17, XD⁺17 | Getting started | 85 Table 1: Table with bucky Table with bucky The MUSICA Acquisition Workstation is configured for a specific patient orientation, either head left (default) or head right. Table 2: Wallstand bucky Wallstand with left loading bucky...
86 | XD 17, XD⁺17 | Getting started Offline image acquisition workflow A prerequisite for the offline image acquisition workflow is that the DR detector is configured to use automatic exposure detection. The offline image acquisition workflow is only intended for DR detectors that are used in a DR Retrofit solution.
Page 87
XD 17, XD⁺17 | Getting started | 87 If the patient data for the first image are known, they can be filled in, otherwise, leave the patient data blank. 2. Add a thumbnail for the offline workflow to the examination.
Page 88
88 | XD 17, XD⁺17 | Getting started The images are downloaded from the DR detector. A message is displayed. Figure 19: A recovery image has been received • The first image is stored on the thumbnail of the original examination.
Page 89
XD 17, XD⁺17 | Getting started | 89 2. Click Transfer Images. The Transfer Images wizard opens. 3. In the Image Overview pane, select the image. The image is displayed in the wizard. 4. Click Continue. 5. In the Worklist window, select the examination with the correct patient data.
90 | XD 17, XD⁺17 | Getting started Guidelines for Pediatric Applications CAUTION: Use special care when imaging patients outside the typical adult size range. Children are more radiosensitive than adults. Reducing dose for radiographic procedures while maintaining acceptable clinical image quality will benefit patients.
Page 91
XD 17, XD⁺17 | Getting started | 91 • Image only the indicated area. • Use the lowest amount of radiation for adequate imaging based on size of the child (reducing tube output -- kVp and mAs, limiting the duration of dynamic imaging).
92 | XD 17, XD⁺17 | Getting started Stopping the DR detector To stop the DR detector: 1. If the DR detector is connected to the System Control Unit via the DR detector cable, turn off the System Control Unit using the power switch.
XD 17, XD⁺17 | Getting started | 93 Automatic exposure detection Depending on the configuration, the DR detector detects X-ray exposure to automatically perform the image acquisition. Before performing the exposure, the DR detector must be ready. Check the status of the DR detector in the DR Detector Switch.
94 | XD 17, XD⁺17 | Getting started Attaching the handle unit with grid CAUTION: To safely attach the handle unit for making exposures using the grid, follow these instructions. 1. Lay down the grid on a flat surface. 2. Lay the DR detector in the grid, bottom edge first, with the tube side facing down (1).
XD 17, XD⁺17 | Getting started | 95 Attaching the handle unit without grid To attach the handle unit for making exposures without using the grid 1. Lay down the handle unit on a flat surface. 2. Lay the DR detector in the handle unit, bottom edge first, with the tube side facing up (1).
XD 17, XD⁺17 | Advanced Operating | 97 Viewing the detector status The status display on the side of the DR detector is turned off by default. The status display is controlled by pressing the power button (or the AP mode button).
98 | XD 17, XD⁺17 | Advanced Operating Battery status Table 3: Battery status No battery is present, or battery charge level is less than 5% The battery is being charged. The bar inside the battery icon repre- sents the current charge level.
XD 17, XD⁺17 | Advanced Operating | 99 Connection status Table 4: Access point mode: the DR detector is acting as access point for wireless network The DR detector is acting as access point for wireless network Table 5: Client mode: the DR detector is connected to a wireless network Preset Identifier (Default "Sy"...
100 | XD 17, XD⁺17 | Advanced Operating Charging a battery There are 4 ways to charge a battery (depending on the available accessories): • Put the detector in an empty slot of the DR detector cradle. • Connect the power adapter with the USB Type-C cable to the DR detector.
XD 17, XD⁺17 | Advanced Operating | 101 Charging the DR detector in the DR detector cradle A DR detector can be charged in either of both slots of the DR detector cradle, or two detectors can be charged simultaneously.
102 | XD 17, XD⁺17 | Advanced Operating Charging the DR detector using the power adapter CAUTION: Charge the battery in an environment within a temperature range of 15 to 35 °C. Not doing so may result in battery leakage, overheating or damage.
XD 17, XD⁺17 | Advanced Operating | 103 Charging the battery using the DR detector cable Connect the DR detector cable to charge the battery that is attached to the DR detector. The battery status can be read from the DR Detector Switch on the MUSICA Acquisition Workstation.
104 | XD 17, XD⁺17 | Advanced Operating Charging a battery in the dual battery charger A battery can be charged in either of both slots of the battery charger, or two batteries can be charged simultaneously. 1. Insert a battery in an empty slot of the battery charger.
XD 17, XD⁺17 | Advanced Operating | 105 Replacing the battery To replace the battery, a flat-head screwdriver is required. For regular use, the other means for charging the DR detector are recommended. Battery replacement may be required if the battery is broken or if there is no time for regular charging and a back-up battery is available.
106 | XD 17, XD⁺17 | Advanced Operating Managing network connections in client mode configuration The DR detector is configured for client mode. Figure 24: Client mode Each workstation has an access point or a System Control Unit. Multiple DR detectors can be connected to a MUSICA Acquisition Workstation simultaneously.
Page 107
XD 17, XD⁺17 | Advanced Operating | 107 4. Press the AP mode button for 1 second to display the SSID name of the access point to which the DR detector is connected. Related Links Configuration on page 14 0434A EN 20210426 1355...
108 | XD 17, XD⁺17 | Advanced Operating Managing network connections in access point mode configuration The DR detector is configured for access point mode. Figure 26: Access point mode Multiple DR detectors can be used on the same MUSICA Acquisition Workstation, but they cannot be connected simultaneously.
XD 17, XD⁺17 | Advanced Operating | 109 Switching between the wireless DR detector and the wireless hospital network The MUSICA Acquisition Workstation can be configured to connect to a wireless DR detector as well as to a wireless hospital network.
Page 110
110 | XD 17, XD⁺17 | Advanced Operating Hospital Network DR Detector 1 DR Detector 2 Figure 28: Available wireless networks • To connect to the hospital network, select name of the hospital network. Wi-Fi The MUSICA Acquisition Workstation is connected to the hospital network to retrieve data from the RIS or to print or archive images.
XD 17, XD⁺17 | Advanced Operating | 111 Switching temporarily to client mode The DR detector is configured in access point mode. The DR detector can be temporarily switched to client mode, to connect to another MUSICA Acquisition Workstation. The detector can be configured upon installation with a single MUSICA Acquisition Workstation (with access point or System Control Unit) to which it will connect in this status.
Page 112
112 | XD 17, XD⁺17 | Advanced Operating Connecting the MUSICA Acquisition Workstation to another DR detector (access point mode) Hospital Network The DR detector is configured in access point mode. WARNING: If the network connection is interrupted, the MUSICA...
XD 17, XD⁺17 | Problem solving | 113 Problem solving Topics: • Artifact in DR Detector images • DR detector not ready for exposure • The MUSICA Acquisition Workstation is connected to the DR detector, but the DR detector is not active (access point mode) •...
114 | XD 17, XD⁺17 | Problem solving Artifact in DR Detector images Details An artifact is visible in the images produced by a DR Detector. Cause Exposure conditions have changed significantly since latest calibration. Brief Solution Perform calibration of the DR Detector.
XD 17, XD⁺17 | Problem solving | 115 DR detector not ready for exposure Details The DR detector is turned on. The DR detector status icon on the DR Detector Switch is not green. Cause The DR detector is not connected to the MUSICA Ac- quisition Workstation.
Page 116
116 | XD 17, XD⁺17 | Problem solving Brief Solution Use the Windows wifi setting to connect the MUSICA Acquisition Workstation to the wireless network of the DR detector. Related Links Viewing the detector status on page 97 Connecting to another MUSICA Acquisition Workstation (client mode)
XD 17, XD⁺17 | Problem solving | 117 The MUSICA Acquisition Workstation is connected to the DR detector, but the DR detector is not active (access point mode) Details The DR detector is configured for access point mode. The Wifi...
118 | XD 17, XD⁺17 | Problem solving Images are not sent to the printer or to the PACS archive Details The exam is closed, but the images are not sent to the printer or to the PACS archive. Cause The MUSICA Acquisition Workstation has not been connected to the hospital network.
Safety is only guaranteed when changes, additions, maintenance or repairs are carried out by an Agfa certified field service engineer. A non certified engineer performing a modification or service intervention on a medical device, acts on his own responsibility and makes the warranty void.
Page 120
120 | XD 17, XD⁺17 | Problem solving Symptom Cause Remedy purchase of consumables, contact your sales representa- tive or local dealer. The battery was In low temperatures the bat- charged or used tery capacity decreases. Use a in low tempera- battery charged in normal tures.
XD 17, XD⁺17 | Technical Data | 121 Technical Data Topics: ⁺ 17 technical data • XD 17, XD • DR detector battery technical data • DR detector cradle technical data • DR detector dual battery charger technical data •...
122 | XD 17, XD⁺17 | Technical Data XD 17, XD⁺17 technical data Commercial name XD 17, XD⁺17 Manufacturer Manufacturer DR Detector Vieworks Co., Ltd. (Gwanyang-dong), 41-3, Burim-ro 170beon- gil, Dongan-gu, Anyang-si, Gyeonggi-do, Ko- Distributor DR Detector Agfa NV Septestraat 27, B-2640 Mortsel - Belgium...
Page 123
XD 17, XD⁺17 | Technical Data | 123 Atmospheric pressure between 700 hPa and 1060 hPa Environmental conditions (during storage and transport) Temperature (ambient) between -15 °C and +55 °C Humidity (non condensing) between 5% and 90% (non-condensing) Atmospheric pressure...
124 | XD 17, XD⁺17 | Technical Data DR detector battery technical data Type of product Rechargeable lithium ion polymer bat- tery pack Part number FXRB-04A Dimensions Dimensions (length x width x 189.0 mm x 89.0 mm x 6.65 mm...
XD 17, XD⁺17 | Technical Data | 125 DR detector cradle technical data Type of product Lithium ion battery pack charger Part number FXRR-01A Simultaneous charging 2 detectors Dimensions Dimensions (width x height x depth) 410.0 mm x 159.0 mm x 168.0 mm...
126 | XD 17, XD⁺17 | Technical Data DR detector dual battery charger technical data Type of product Lithium ion battery pack charger Part number FXRC-04A Simultaneous charging 2 batteries Dimensions Dimensions (length ✕ width ✕ 304.0 mm ✕ 230.0 mm ✕ 15.5 mm...
XD 17, XD⁺17 | Technical Data | 127 System Control Unit Part number FXRS-03A Rated power supply AC100 to 240V, 50/60 ㎐, Max. 2.0-0.8A (input) Rated power supply DC +24V 3.25A, 78W (output) Wireless connection IEEE 802.11n (2.4 GHz/5 GHz)
128 | XD 17, XD⁺17 | Technical Data Mini System Control Unit Part number FXRS-04A Rated power supply DC +24V 2A max (input) Wireless connection IEEE 802.11n (2.4 GHz/5 GHz) Dimensions (width x height x depth) 210 mm x 170 mm x 45 mm...
XD 17, XD⁺17 | Technical Data | 129 System Control Unit Lite Part number FXRP-02A DC +24V 1A max Rated power supply (input) Dimensions (width x height x depth) 109 mm x 108 mm x 29.5 mm Weight 0.33 kg...
XD 17, XD⁺17 | Remarks for HF-emission and immunity | 131 EMC (Electromagnetic Compatibility) Statements WARNING: This device has been tested for EMI/EMC compliance, but interference can still occur in an electromagnetically noisy location. Attempt to maintain a suitable distance between electrical devices to prevent malfunction.
132 | XD 17, XD⁺17 | Remarks for HF-emission and immunity Electromagnetic emissions This device has been tested for a normal hospital environment as described below. The user of the device should ensure that it is used in such an environment.
XD 17, XD⁺17 | Remarks for HF-emission and immunity | 133 Electromagnetic immunity This DR Detector is intended for operation in the electromagnetic environment given below. The user of the DR Detector should ensure that it is used in such an environment.
Page 134
134 | XD 17, XD⁺17 | Remarks for HF-emission and immunity 100% reduc- 100% reduc- terruptions or a bat- tion for tion for tery. 250/300 cycles 250/300 cy- (5 sec.) cles (5 sec.) Magnetic field at 30 A/m 30 A/m...
Page 135
XD 17, XD⁺17 | Remarks for HF-emission and immunity | 135 Note: These Guidelines may not apply to all situations. The dis- persion of electromagnetic waves is influenced by absorption and reflections from buildings, objects and people. WARNING: Field strengths from fixed transmitters, such as base stations...
136 | XD 17, XD⁺17 | Remarks for HF-emission and immunity For U.S.A. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: • This device may not cause harmful interference. • This device must accept any interference received, including interference that may cause undesired operation.