Detailed Instrument Functions
CAL: This function indicates the date of the last calibration (format DD/MM/YY); long press the right
button to display.
The bAtt function indicates the battery charge status (from 10% to 90%).
N.B. for recharging see page 23.
NEt function indicates the reception status of the GSM/GPS system.
The Point function allows you to enter the position number after taking a measurement. The position
number is described in the Quick Guide supplied with the instrument. By default, this function is
deactivated, to activate it long press on the right button, then a short press to confirm.
The dynA function allows the instrument to be used as a "dynamometer" (e.g., for measuring the pushing
force of safety doors). To take a measurement, long press the right button. To exit this function, press the
right button again.
The UPdAtE function allows the firmware of the instrument internal processor to be updated via web.
To start the update, long press the right button. After a few minutes the instrument will switch off.
N.B. If the existing firmware has already been updated, the display will show the number of the currently
saved version. Before doing this, make sure that the battery is charged (at least 40%). The update will not
start if the USB cable is connected.
Warning Messages
If the message bAtt is displayed when turning on, the battery charge is at a minimum (10%).
CAL: This warning is displayed one month before the calibration due date. When more than one year has
passed, the warning will be displayed each time the instrument is turned on.
If the message FULL is displayed when turning on, all memory locations are occupied, you need to
download or delete the saved measurements to make new ones.
The message Err 1 indicates that the USB cable is connected but is not required for the current function.
The message Err 2 indicates that the instrument does not communicate with the PC. (See also page 15)
The message Err 3 indicates that communication with the server has failed.