34 | WS-Motorized-T-001 | Radiographic Wall Stand
Automatic centering on the wall stand with the X-ray tube not perpendicular
to the bucky, will cause the laser lines not to coincide with the center marks on
the front panel of the bucky, because there is a distance between the front
panel and the cassette or detector.
Topics:
•
Positioning the Radiographic Wall Stand
•
Tracking the radiographic wall stand
•
Radiographic Wall Stand Accessories
3255F EN 20210312 1125
CAUTION:
The format indications on the front panel show the format of the
cassette or detector. Take into account that the actual area for
imaging is smaller than indicated. The image of the exposed
object is slightly magnified because there is a distance between
the front panel and the cassette or detector. The sensitive area of
the cassette or detector may be slightly smaller than the
indicated area. Check the technical data of the cassette or
detector for exact values.