AGFA WS-Motorized-T-001 User Manual page 62

Table of Contents

Advertisement

62 | WS-Motorized-T-001 | Product Information
TP_T;
StC
Master ME/ES Signal
CAN Signal
Bucky D. exp. / Bucky se-
lect T
Float-GND
230V Power Supply
PE Cable
AEC
Sync 1/2 Cable (Varian)
Image Data Line
Cable CR_PC_NIC - WS
Bucky (direct wire)
TP_WS;
StC
ME/ES Signal
Positioning Control
Box
3255F EN 20210312 1125
type;
maximum length (m)
3 x AWG22;
20
Standard RS-232 cable
(9 Pin D-SUB);
20
7 x AWG20;
20
1x AWG12;
20
3x AWG18;
20
1x AWG10;
20
CAT 5e SF/UTP or F/
UTP;
20
8 x AWG26;
16
CAT6;
40
type;
maximum length
(m)
2 x AWG22;
20
4 x AWG20;
20
remark
Basic wiring
Basic wiring / Extension
not allowed
Basic wiring
Basic wiring
Basic wiring
Basic wiring
Basic wiring
Customized wiring /
PIN 9 NC
Customized wiring / di-
rect wire - Point to
Point / Extension not al-
lowed
remark
Basic wiring for DR 600 WS
Basic wiring for DR400-600 WS
2 more Contorl Boxes can be
added in the system