GE B20 User Manual page 39

Patient monitor
Hide thumbs Also See for B20:
Table of Contents

Advertisement

Changing alarm options
Monitor Setup
1. Press
key.
2. Select Install/Service and enter the password.
3. Select Installation - Alarm Options.
Show Limits: Select YES to show alarm limits in digit fields.
YES is the default.
Show Audio ON/OFF: Select YES to enable alarm audio off.
Select NO (default) disables audio off option in the Audio ON/
OFF menu in Alarms Setup.
Latching Alarms: Select YES to keep alarm messages on
Audio pause
screen until
pressed. NO is the default.
Breakthrough Alarm: Select Off to close breakthrough
alarms feature, select Pedi to turn on breakthrough feature
for new fatal alarms, select Adult to turn on breakthrough
feature for new fatal alarms, excluding Brady.
Min Alarm Volume: Adjust minimum volume of audible
alarm tone.
Reminder Volume: Adjust volume of audible alarm reminder
tone.
Alarm Tones: Select ISO, ISO2, General or IEC tones.
Remote Control: Select active to enable alarm remote
control by central station. Inactive is the default.
Alarm Reset
key or
key is
Changing the printer settings
Monitor Setup
1. Press
2. Select Install/Service and enter the password.
3. Select Installation - Printer.
Snapshot Printout: Select 12.5 or 25 mm/s
Printer Connection: Select Net 1 to Net 16/None (default)
Paper Size: Select A4 or Letter
NOTE: Network printer only.
Other printer settings can be changed in the Record & Print menu.
For more information about the printing function, see the
and recording"
chapter.
Changing the parameter settings
Monitor Setup
1. Press
2. Select Install/Service and enter the password.
3. Select Installation - Monitor Settings - Parameter Settings.
CO2 Numbers: Set CO
MAC Type: Set MAC type MAC or MACage (B40 only)
35
key.
key.
humidity compensation Dry or Wet.
2
"Printing