GE B20 User Manual page 127

Patient monitor
Hide thumbs Also See for B20:
Table of Contents

Advertisement

Printing
Printing is not possible:
Check the printer setting through
Connection.
Check that the printer is connected to the network.
Check the network cable.
Recording
Recording is not possible:
Check the Central recorder if you are recording through
network.
Non-invasive blood pressure
Measurement does not work or values seem unstable:
Check that cuff tubings are not bent, stretched, compressed
or loose.
When using hoses without identification, make sure that you
have selected the inflation limits in the NIBP Setup menu, see
"Non-invasive blood pressure
Prevent motion artifacts.
Use cuffs of correct size.
Print/Record
- Printer
(NIBP)."
Pulse oximetry
SpO
signal is poor:
2
Check the sensor and sensor positioning.
If in GE SpO
configuration, change the SpO2 Response
2
(averaging time) to Normal.
Note that skin pigment causes differences.
Make sure that the patient is not moving.
Temperature
Measurement fails:
Check that the probe is properly connected to the probe
adapter.
Check that you are using the correct probe for the
anatomical location being monitored.
Use a probe that is compatible with your system.
Try using a known good probe in case the sensor is
damaged.
Check the patient connection.
123