Printing
Printing is not possible:
•
Check the printer setting through
Connection.
•
Check that the printer is connected to the network.
•
Check the network cable.
Recording
Recording is not possible:
•
Check the Central recorder if you are recording through
network.
Non-invasive blood pressure
Measurement does not work or values seem unstable:
•
Check that cuff tubings are not bent, stretched, compressed
or loose.
•
When using hoses without identification, make sure that you
have selected the inflation limits in the NIBP Setup menu, see
"Non-invasive blood pressure
•
Prevent motion artifacts.
•
Use cuffs of correct size.
Print/Record
- Printer
(NIBP)."
Pulse oximetry
SpO
signal is poor:
2
•
Check the sensor and sensor positioning.
•
If in GE SpO
configuration, change the SpO2 Response
2
(averaging time) to Normal.
•
Note that skin pigment causes differences.
•
Make sure that the patient is not moving.
Temperature
Measurement fails:
•
Check that the probe is properly connected to the probe
adapter.
•
Check that you are using the correct probe for the
anatomical location being monitored.
•
Use a probe that is compatible with your system.
•
Try using a known good probe in case the sensor is
damaged.
•
Check the patient connection.
123