Setting IBP alarms
IBP
1. Press the
key.
2. Select IBP1 Setup - IBP1 Alarm and IBP2 Setup - IBP2 Alarm.
3. Select Adjust Limits to set up the limits.
4. Select Sys Alarm, Dia Alarm or Mean Alarm to OFF/ON the
limits.
5. You can also select Alarms Setup to adjust IBP or related alarms.
Troubleshooting
Readings seem unstable:
Make sure that there are no air bubbles in the transducer
system.
Flush and zero.
Place the transducer on the patient's mid-heart level and
zero.
Invasive blood pressure waveform is displayed but no numeric
values are displayed:
Zero the channel. Invasive blood pressure numerical values
are displayed only for successfully zeroed channels.
99