Long measurement time
−
NIBP: reduce patient's motion.
•
Low gas sample flow
−
Check the patient status.
−
Change the sampling line or water trap.
•
Mark x
−
A snapshot is created manually.
•
Memory ERROR
−
Contact authorized service personnel.
•
Mode Data Reset
−
Contact authorized service personnel.
•
Module power supply overload
−
Check the patient status.
−
Contact authorized service personnel.
•
Multiple agents present
−
Check if the anesthetic agents are more than one.
−
Check the patient status.
•
Network: x
−
Network is connected.
•
Network down: xxx (xxx = network name)
−
Try to re-establish the connection.
−
Contact authorized service personnel.
•
Network recording...
126
−
Network recorder has been started. Please wait until the
recording is finished.
•
NIBP call service error
−
NIBP: Contact authorized service personnel.
•
NIBP manual
−
Check the cuff and cuff hose.
•
NIBP cuff overpressure
−
Check NIBP cuff hose and tubes.
−
Restart measurement.
•
NIBP cuff loose
−
Check the cuff and cuff hose.
•
NIBP cuff occlusion
−
Check the tubes and hose.
•
No battery backup
−
Replace the batteries, see
•
No Entropy sensor
−
Check connection between Entropy sensor and cable.
•
No Transducer
−
IBP: Connect the transducer or the cable.
•
No SpO2 probe
−
Check connection between the SpO
•
No SpO2 pulse
−
Try other measuring sites.
"Replacing the
batteries."
sensor and module.
2