Modifying other adjustable screen feature
Display brightness
Monitor Setup
1. Press
and select Display brightness.
2. Select from 10 to 100 %.
Sweep speeds
You can change the speed of the waveforms on the screen.
Hemodynamic and EEG waveform sweep speed options are 12.5,
25 and 50 mm/s.
Other parameters, waveform sweep speed options are Fast
(6.25 mm/s) and Slow (0.625 mm/s).
NOTE: If the sweep speed is set to Slow, the sweep speed is one tenth
of a full screen sweep. Slow sweep speed waveforms show
amplitude changes better than sweep speeds set to Fast.
Monitor Setup
1. Press the
2. Select Sweep Speeds.
3. Select the parameter and adjust the value.
key.
Parameter colors
Parameter color options include: yellow, white, green, red, blue,
orange and violet. To change the color perform the following steps:
Monitor Setup
1. Press the
2. Select Install/Service and enter the password.
3. Select Colors.
4. Select the desired parameter and the color.
Displaying pulse rate
Combined heart rate and pulse rate can be displayed next to the
ECG waveform. The current HR source is displayed in larger font. A
flashing heart symbol is displayed next to the HR value.
ECG
1. Press the
key.
2. Select ECG Setup.
3. Select Display with HR to PR, PVC or None.
31