Adjusting the no breath time
If the monitor has the NeoResp license, in the neonatal mode, the
impedance respiration can set up the detect time for alarm "No
Breath".
Others
1. Press the
key
2. Select Resp Setup - No Breath Time
The default detection time is 15s.
Adjusting respiration alarms
Others
1. Press the
key.
2. Select Resp Setup - Resp Rate Alarm.
•
To set up the limits, select Adjust Limits.
•
To OFF/ON the limits, select RR Alarm.
3. You can also select Alarms Setup to adjust RESP or related
alarms.
Troubleshooting
•
Measurement fails:
−
Check the electrode quality and positioning.
−
Adjust the detection limits. During ventilator supported
breathing, the respiration calculation may count only
ventilator-produced inspirations and expirations.
−
Other electrical devices may interfere with the measurement.