RTC (Real Time Control) Operation

This centrifuge contains an internal clock allowing you to operate the centrifuge between a set start time and end time� This function
for operating the centrifuge between set time intervals is called Real Time Control (RTC)�
The RTC operation procedures are described below using examples�
Example: To operate the centrifuge under the following run parameters, beginning with rotor installation on October 24 and removing
the samples around 9:30 the next morning�
1�
Rotor: T-8100
2�
Speed: 100,000 rpm
3�
Separation time: 3 hours
4�
Control temperature: 4 °C
5�
Acceleration mode: 9
6�
Deceleration mode: 9
In this example, set the above run parameters (1) to (6), then set the RTC operation start time to 6:30 on October 25, and start the
centrifuge�
(Note that the same operation will be performed if you set an end time of 9:30 instead of the start time (6:30)�)