Rtc (Real-Time Control) Operation - Thermo Scientific Sorvall WX+ Instruction Manual

Ultracentrifuge
Table of Contents

Advertisement

3-3-4 RTC (Real Time Control) Operation
This centrifuge contains an internal clock allowing you to operate the centrifuge between a set start
time and end time. This function for operating the centrifuge between set time intervals is called Real
Time Control (RTC).
The RTC operation procedures are described using examples below.
Example: To operate the centrifuge under the following run parameters starting from rotor installation
on October 24 and removing the samples around 9:30 the next morning.
(1) Rotor: T-8100
(2) Speed: 100,000 rpm
(3) Separation time: 3 hours
(4) Control temperature: 4 °C
(5) Acceleration mode: 9
(6) Deceleration mode: 9
April 1 PM 4:30
October 24
In this example, set above run parameters (1) to (6), then set the RTC operation start time to 6:30
October 25 and start the centrifuge.
(Note that setting an end time of 9:30 instead of the start time (6:30) will perform the same
operation.)
[START] button
START button
ON
October 25
Fig. 3-3-4 RTC Operation Example
3-54
3-3-4 RTC (Real Time Control) Operation
3 h
AM 6:30
Auto start
End time
Around 9:34
Stop

Hide quick links: