Strain relieving the cables

If the workstation table (option) is used with the Opdima system, the cables from the
workstation components and from the MAMMOMAT must be strain relieved by cable
clamps at the back of the MO unit shelf according to the figure below.
To ensure that the cables have enough length for the vertical adjustment, let the
workstation table be in its highest possible position when strain relieving the cables.
Use cable ties to arrange the cables in an organized way and make sure that the cables
can run without a risk of getting caught.
Biopsy
controller
Fig. 26

Functional test (workstation table)

Before the installed workstation table (option) is handed over to the customer, check that
the workstation table can be vertically adjusted between the highest and lowest possible
positions properly and safety.