Emerson Rosemount 1057 Quick Start Manual page 15

Triple channel transmitter
Hide thumbs Also See for Rosemount 1057:
Table of Contents

Advertisement

February 2022
Conduit openings accept 0.5 in (13 mm) conduit fittings or PG13.5 cable
glands. To keep the case watertight, block unused openings with Type 4X or
IP66 conduit plugs.
NOTICE
Use watertight fittings and hubs that comply with your requirements.
Connect the conduit hub to the conduit before attaching the fitting to the
transmitter.
3.3
Power, output, and sensor connections
3.3.1
Power wiring
Emerson offers two power supplies for the Rosemount 1057:
1. 24 Vdc (20-30 V) power supply (02 ordering code)
2. 85-265 Vac switching power supply (03 ordering code)
AC mains (115 or 230 V) leads and 24 Vdc leads are wired to the power
supply board, which is mounted vertically on the left side of the main
enclosure cavity. Each lead location is marked clearly on the power supply
board. Wire the power leads to the power supply board using the lead
markings on the board.
The grounding plate is connected to the earth terminal of power supply
input connector TB1 on the 01 (115/230 Vac) and 03 (85-265 Vac) power
supplies. The green screws on the grounding plate are intended for
connection to some sensors to minimize radio frequency interference. The
green screws are not intended to be used for safety purposes.
Quick Start Guide
Quick Start Guide
15