750000 AlternAting Pressure / low Air loss MAttress systeM user MAnuAl important: Do not operate the Mattress system without first reading and understanding this manual! Save this manual for future use. 750000-INS-LAB-RevA08...
IntroduCtIon Use this manual for set-up and operation of the Alternating Pressure / Low Air Loss Mattress System. Read all instruc- tions before using the mattress system. Save this manual for future reference. Intended use of thIs devICe The intended use of the Alternating Pressure / Low Air Loss Mattress System is: •...
danger to reduce the risk of electrocution: DANGER: Always unplug this product immediately after using. DANGER: Do not use this product while bathing. DANGER: Do not place or store this product where it can fall or be pulled into a tub or sink. DANGER: Do not place this product in or drop into water or other liquid.
WARNING: Keep the cord away from heated surfaces. WARNING: Never block the air openings of this product or place it on a soft surface, such as a bed or couch, where the openings may be blocked. Keep the air opening free of lint, hair, and other similar particles.
equIpment symbols Attention! Consult accompanying documents (this manual) Indicates that this product is according to the degree of protection against electric shock for type BF equipment Functional earth (for UL only) Class II (for UL only) Disposal of Electrical & Electronic Equipment (WEEE): Do not treat this product as household waste.
produCt desCrIptIon pump and mattress system foot end symbol CPR valves mattress pump mattress-pump connectors pump and mattress system pump: front quick connector 750000-INS-LAB-RevA08...
pump: rear pump: top panel static function switch main power switch low pressure indicator normal pressure indicator pressure adjustment knob Pressure adjust knob: The pressure adjust knob controls the air pressure output. Turn it clockwise to increase output pressure (firmer); turn it counter-clockwise to de- crease air output pressure (softer).
Static function switch: Press the static function switch to suspend alternating function, if needed; the pressure inside the air tubes will then be adjusted to identical pressure. Press the static function switch again to change back to al- ternating mode. The static function switch toggles between static and alternating modes.
Page 10
place the pump on a flat surface easily accessible to the user, care- giver, and/or doctor. s Caution: Place the device in a po- sition where the user, caregiver, and/or doctor can access it easily. 5. Connect the mattress-pump air hose quick connector;...
operatIon 1. Turn the power switch to the ON position. 2. Adjust the pressure adjust knob to the maxi- mum position. Note: Set the pressure adjust knob to the maximum position for the quickest inflation at set-up; the mattress can then be adjusted to the desired pressure. 3. The pump will begin delivering air into the mattress. While the mat- tress is filling, the low pressure indicator (yellow LED) will illuminate.
the patient's buttock area; your fingers should just fit. If the fiit is too loose, decrease the setting. If your fingers do not fit well, increase the setting two notches firmer. Wait ten minutes and then check again. You should only have to ad- just this setting one time per patient.
s Caution: Do not use corrosive or powdered cleansers to clean the pump. s Caution: Do not immerse or soak the pump. s Caution: Do not clean with phenolic products. Mattress: Wipe the mattress unit with a clean, damp cloth and mild detergent.
Air filter replacement 1. Remove the air filter plate locat- ed at rear of pump (filter location shown in picture at right). 2. Remove the filter. The filter is reusable, if not torn, and can be washed gently with a mild de- tergent and water.
Specifications continued Pump Specification Environment Temperature Operation: 50°F to 104°F (10°C to 40°C) Storage: 5°F to 122°F ( -15°C to 50°C) Shipping: 5°F to 158°F ( -15°C to 70°C) Humidity Operation: 10% to 90% non-condensing Storage: 10% to 90% non-condensing Shipping: 10% to 90% non-condensing Classification Class II, Type BF, IPX0...