Agilent 81950A Compact TLS module, First Edition
None
sour1:freq:grid 50e9
Only available in "Grid Mode"
[:SOURce[n]][:CHANnel[m]]:FREQuency:GRID?
[:SOURce[n]][:CHANnel[m]]:FREQuency:GRID?<wsp>[MIN|MAX|DEF]
Returns the grid spacing for the frequency grid used by the instrument.
Optional:
MIN - minimum programmable value
MAX - maximum programmable value
DEF - default value
The current grid spacing value or the minimum, maximum, or default grid spac-
ing value in Hz.
sour1:freq:grid?
+5.00000000E+010
[:SOURce[n]][:CHANnel[m]]:FREQuency:CHANnel
[:SOURce[n]][:CHANnel[m]]:FREQuency:CHANnel<wsp>
<value>|MIN|MAX|DEF
Sets the laser's grid channel.
Any value resulting in a valid frequency, refer to the chapter "Output Frequen-
cy" on page 33.
Also allowed:
MIN - minimum programmable value
MAX - maximum programmable value
DEF - default value
None
sour1:freq:chan -20
Only available in "Grid Mode".
The minimum, maximum, and default value depend on the current reference fre-
quency and grid spacing values.
Programming Instructions