Programming Instructions
Description
Parameters
Response
Example
Affects
Notes
Command
Syntax
Description
Parameters
Response
Example
Affects
Notes
Command
Syntax
Description
Parameters
62
Set the reference frequency for the frequency grid used by the instrument.
Any value in the specified range (see appropriate User's Guide)
Also allowed:
MIN - minimum programmable value
MAX - maximum programmable value
DEF - default value
None
sour1:freq:ref 193.1THz
Only available in "Grid Mode"
[:SOURce[n]][:CHANnel[m]]:FREQuency:REFerence?
[:SOURce[n]][:CHANnel[m]]:FREQuency:REFerence?<wsp>[MIN|MAX|DEF]
Returns the reference frequency for the frequency grid used by the instrument.
Optional:
MIN - minimum programmable value
MAX - maximum programmable value
DEF - default value
The current reference frequency value or the minimum, maximum, or default
reference frequency value in Hz.
sour1:freq:ref?
+1.93100000E+014
[:SOURce[n]][:CHANnel[m]]:FREQuency:GRID
[:SOURce[n]][:CHANnel[m]]:FREQuency:GRID<wsp>
<value>[HZ|KHZ|MHZ|GHZ|THZ]|MIN|MAX|DEF
Set the grid spacing for the frequency grid used by the instrument.
Any value in the specified range (see appropriate User's Guide)
Also allowed:
MIN - minimum programmable value
MAX - maximum programmable value
DEF - default value
Agilent 81950A Compact TLS module, First Edition