Bender isoMED427 Series Manual

Bender isoMED427 Series Manual

Insulation monitoring device for medical applications
Hide thumbs Also See for isoMED427 Series:

Advertisement

Quick Links

isoMED427x-(PT)
Insulation monitoring device for medical applications
isoMED427x-(PT)_D00440_00_M_XXDE
Manual EN

Advertisement

Table of Contents
loading

Summary of Contents for Bender isoMED427 Series

  • Page 1 isoMED427x-(PT) Insulation monitoring device for medical applications isoMED427x-(PT)_D00440_00_M_XXDE Manual EN...
  • Page 2 Service and support for Bender products First level support Technical support Carl-Benz-Strasse 8 • 35305 Grünberg • Germany Telephone: +49 6401 807-760 0700BenderHelp * Fax: +49 6401 807-629 E-mail: support@bender-service.de Available from 7.00 a.m. to 8.00 p.m. on 365 days of the year (CET/UTC+1) * Landline German Telekom: Mon-Fri 09:00 a.m.
  • Page 3: Table Of Contents

    Signs and symbols............................5 Training courses and seminars........................ 6 Delivery conditions............................6 Inspection, transport and storage......................6 Warranty and liability..........................6 Disposal of Bender devices........................7 Safety................................. 7 Device-speci c safety information..............8 Intended use..............................8 Functional description..................9 Alarm and operating messages for each BMS channel............... 10 Mounting and connection................
  • Page 4 Table of contents Password protection..........................21 Deactivating the CT monitoring function..................22 6.10 Resetting the device to factory settings................... 22 6.11 Querying device information.........................23 Technical data isoMED427(P)-(PT)..............24 Factory settings isoMED427x-(PT)....................... 29 Device overview............................30 Recommended device combinations....................31 Standards and certi cations........................31 isoMED427x-(PT)_D00440_00_M_XXDE...
  • Page 5: General Instructions

    This manual is intended for quali ed personnel working in electrical engineering and electronics! Part of the device documentation in addition to this manual is the enclosed supplement " Safety instructions for Bender products". Read the operating manual before mounting, connecting and commissioning the device.
  • Page 6: Training Courses And Seminars

    Regular face-to-face or online seminars for customers and other interested parties: www.bender.de -> Fachwissen -> Seminare. Delivery conditions The conditions of sale and delivery set out by Bender GmbH & Co. KG apply. These can be obtained in printed or electronic format. The following applies to software products: "Softwareklausel zur Überlassung von Standard- Software als Teil von Lieferungen,...
  • Page 7: Disposal Of Bender Devices

    Disposal of Bender devices Abide by the national regulations and laws governing the disposal of this device. For more information on the disposal of Bender devices, refer to www.bender.de -> Service & Support. Safety If the device is used outside the Federal Republic of Germany, the applicable local standards and regulations must be complied with.
  • Page 8: Device-Speci C Safety Information

    Device-speci c safety information Device-speci c safety information Risk of property damage due to unprofessional installation! CAUTION If more than one insulation monitoring device is connected to a conductively connected system, the system may be damaged. If several devices are connected, the device does not work and does not signal insulation faults.
  • Page 9: Functional Description

    In the event of a malfunction, relay K1 switches and all three LEDs ash. The display shows an error code. E.01 Protective conductor connection faulty, no low-resistance connection between E and KE. E.03 Measuring current transformer interruption E.04 Measuring current transformer short circuit E.05…E.xx Internal device error, contact Bender service. isoMED427x-(PT)_D00440_00_M_XXDE...
  • Page 10: Alarm And Operating Messages For Each Bms Channel

    Functional description Password protection If the password protection has been enabled, settings can only be made after entering the correct password. Factory settings FAC After restoring the factory settings, all settings previously changed are reset to the state upon delivery. Monitoring of the IT system transformer For temperature monitoring, the isoMED427P-PT device evaluates the resistance value of the temperature sensor.
  • Page 11 isoMED427x-(PT) Operating messages BMS channel Description BMS channel Description Currently measured insulation resistance R F Currently measured load current in % Current transformer temperature in °C (isoMED427P-PT only) isoMED427x-(PT)_D00440_00_M_XXDE...
  • Page 12: Mounting And Connection

    Mounting and connection Mounting and connection Risk of fatal injury due to electric shock! DANGER Touching live parts of the system carries the risk of an electric shock. • Before installing and connecting the device, make sure that the installation has been de-energised.
  • Page 13: Wiring Diagram

    isoMED427x-(PT) Wiring diagram Risk of electrocution due to electric shock! DANGER If terminals L1, L2 of the device are connected to an energised IT system, terminals E and KE must not be disconnected from the protective conductor (PE). Connect the device according to the wiring diagram. The lines to KE and E must be routed separately! The BMS bus must be terminated at both ends! STW2...
  • Page 14: Commissioning

    Mounting and connection Commissioning Prior to commissioning, check proper connection of the ISOMETER®. Perform a functional test using a genuine earth fault, e.g. via a suitable resistance. isoMED427x-(PT)_D00440_00_M_XXDE...
  • Page 15: Operating Elements

    isoMED427x-(PT) Operating elements Device front Element Function Power on LED Alarm LED 1 is lit: Below response value R Alarm LED 2 is lit: Response value % I or/and °C exceeded "T": Start a self test (2 s) : Menu item up / increase value : Menu item down / decrease value Start menu mode (2 s) Enter button:...
  • Page 16: Display Elements Isomed427X-(Pt)

    Operating elements Display elements isoMED427x-(PT) After activating the test function (T), all available display elements are shown for 2 seconds. < | - | > Identi er for response values 7-segment display Display of the values Upper row R : Insulation resistance I : Fault current .
  • Page 17: Alarm Indication And Eds Activity

    isoMED427x-(PT) Alarm indication and EDS activity Display examples: > > > isoMED427(P) C ° > isoMED427P-PT If di erent alarms occur simultaneously, the display changes at 3-second intervals. If several errors occur, the respective error codes are displayed alternately. isoMED427x-(PT)_D00440_00_M_XXDE...
  • Page 18: Parameter Settings

    Parameter settings Parameter settings An example is given below on how to change the alarm response value R (< R). Proceed as follows: Press and hold the MENU/Enter button for 2 s. The ashing short symbol < R appears on the display.
  • Page 19: Menu Overview

    isoMED427x-(PT) Menu overview > < 1.5 s < R [k ] > > . . . 1.5 s > I [mA] 1.5 s > °C [°C] < t > 5 Min. < > 1.5 s < R [k ] > I [mA] >...
  • Page 20: Setting The Response Value (

    Parameter settings Setting the response value R (<R) Use this setting to de ne the insulation value below which an alarm is signalled. > > > > Setting the response value for load current monitoring > > > > > Setting the response value overtemperature ( >°C) (isoMED427P-PT) C °...
  • Page 21: Activating Automatic Insulation Fault Location

    isoMED427x-(PT) Activating automatic insulation fault location > Password protection This menu can be used to enable the password protection, change the password or to disable the password protection. The password can only be changed if the password protection has been enabled before. a) Enabling the password protection >...
  • Page 22: Deactivating The Ct Monitoring Function

    Parameter settings c) Disabling the password protection > Deactivating the CT monitoring function > 6.10 Resetting the device to factory settings > isoMED427x-(PT)_D00440_00_M_XXDE...
  • Page 23: Querying Device Information

    isoMED427x-(PT) 6.11 Querying device information This function is used to query the software (1.xx) version. After activating this function, the data will be displayed as scrolling text. After completing the routine, individual data sections can be selected using the arrow-up/arrow-down buttons. >...
  • Page 24: Technical Data Isomed427(P)-(Pt)

    Technical data isoMED427(P)-(PT) Technical data isoMED427(P)-(PT) Insulation coordination acc. to IEC 60664-1/-3 De nitions Measuring circuit (IC1) L1, L2 Control circuit (IC2) E, KE, Z, Z/k, l, T1, T2, A, B Output circuit (IC3) 11, 12, 14 Rated voltage 250 V Overvoltage category Operating altitude <...
  • Page 25 isoMED427x-(PT) Monitored IT system Nominal system voltage U n 70…264 V Nominal frequency f n 47…63 Hz Insulation monitoring acc. to IEC 61557-8: 2007-01 Response value R an 50…500 k Relative uncertainty ±10 % Hysteresis 25 % Response time t an at R F = 0.5 x R an and C e = 0.5 F ≤...
  • Page 26 Technical data isoMED427(P)-(PT) Response time, overload (50 % to 120 %) < 5 s Response time, CT monitoring at restart, test or every 1 h Temperature monitoring isoMED427x Sensor PTC resistors acc. to DIN 44081 (max. 6 in series) Response value Release value 1.6 k Relative uncertainty...
  • Page 27 isoMED427x-(PT) Interface Interface/protocol RS-485/BMS Baud rate 9.6 kBit/s Cable length ≤ 1 200 m Cable: twisted pair, one end of shield connected to PE recommended J-Y(St)Y min. n x 2 x 0.8 120 Ω (0.25 W), internal, switchable Terminating resistor Device address, BMS bus 2…90 Switching elements...
  • Page 28 Technical data isoMED427(P)-(PT) Classi cation of mechanical conditions acc. to IEC 60721 Stationary use (IEC 60721-3-3) 3M11 Transport (IEC 60721-3-2) Long-term storage (IEC 60721-3-1) 1M12 Connection Connection type Push-wire terminals Nominal current ≤ 10 A Stripping length 10 mm Opening force 50 N Test opening, diameter 2.1 mm...
  • Page 29: Factory Settings Isomed427X-(Pt)

    isoMED427x-(PT) Other Operating mode Continuous operation Position of normal use Degree of protection, built-in components (DIN EN 60529) IP30 Degree of protection, built-in components (DIN EN 60529) IP20 Enclosure material Polycarbonate Flammability class UL94V-0 DIN rail mounting IEC 60715 Screw mounting 2 x M4 Software versions D643 V1.0x (isoMED427-2)
  • Page 30: Device Overview

    Technical data isoMED427(P)-(PT) Device overview Model Supply voltage U s Article number isoMED427-2 AC 70...264 V; 47…63 Hz B72075306 B92075306 isoMED427P-2 B72075301 B92075301 isoMED427P-PT B72075307 B92075307 Push-wire terminal version Screw-type terminal version Only this device has a Lloyds Register approval Accessories Description Article number...
  • Page 31: Recommended Device Combinations

    isoMED427x-(PT) Recommended device combinations Device combination Insulation Insulation Alarm indicator and monitoring device fault locator test combination isoMED427P(-PT) = S EDS461-L = S MK2430, CP9xx, CP305 = M EDS461-D = S MK2007 = S isoMED427P(-PT) = S EDS151 = S MK2430, CP9xx, CP305 = M MK2007 = S S = Slave, M = Master...
  • Page 36 Nachdruck und Vervielfältigung Reprinting and duplicating nur mit Genehmigung des Herausgebers. only with permission of the publisher. Bender GmbH & Co. KG Bender GmbH & Co. KG Postfach 1161 • 35301 Grünberg • Deutschland PO Box 1161 • 35301 Grünberg • Germany Londorfer Str.

This manual is also suitable for:

Isomed427 pt series

Table of Contents