Technical Data - Dräger Prana W Instructions For Use Manual

Bed head units for medical applications
Table of Contents

Advertisement

Technical Data

Size
Rows of outlets
Length
Any length upto 3 mtrs.
Outlet spacing
At least 200mm between gas terminals units (for flammable or oxidizing gases) and electrical
outlets.
Outlets
Dräger DIN and BS Gas terminal units for medical gases and vacuum, Indian Standard Institute
(ISI) certified Electrical sockets (max. 16 A), data and telephone sockets or Cut outs for parts
provided by customers, e.g. gas outlets, electrical socket, nurse call, etc.
Flush Wall
Brick Wall - Use feeding bracket.
Mount
Ply Wall- Use wood screws for mounting
Gypsum Board Wall- Use mounting angles for mounting
Provision for entry of supply lines from the back side/ Left or Right side of Flush wall supply
unit.
Installation shall be carried out as per Installation Instruction
Profile colours
4 standard RAL colours to choose from-Signal Yellow (RAL 1003),White Aluminium (RAL 9006) ,
Traffic White (RAL 9016),Pastel Blue (RAL 5024), other RAL colours available on request.
Prana W (Single Duct)
1
Prana W (Double Duct)
2
21
Prana W (Triple Duct )
3