Troubleshooting, Service, And Repair Procedures; General Troubleshooting - CAIRE Companion 5 Provider Technical Manual

Oxygen concentrator
Hide thumbs Also See for Companion 5:
Table of Contents

Advertisement

Troubleshooting, Service,
and Repair Procedures
CAUTION The Companion 5 contains electrostatic
sensitive components. Do not open or handle
except at a static free workstation. Do not
remove cover with-out ESD protection.

General Troubleshooting

Before reviewing the troubleshooting chart, the following
steps may be useful to isolate any malfunctions:
1. Turn on the concentrator. If the unit does not turn on, refer
to the troubleshooting chart.
2. Allow the unit to warm up for approximately 10 minutes and
check oxygen concentration.
3. Perform the flow rate test. Verify flows are within
acceptable range per Table 8.
4. Make sure the unit is cycling properly by observing the flow
meter ball is stable (flow meter ball does not move up and
down more than ¼ liter.)
5. Place your thumb over outlet of unit. The flowmeter ball
should drop to the bottom of the flowmeter. If the ball does
not drop completely to the bottom, there is a leak present
between the top of the flowmeter and the outlet of the unit.
6. If concentrator is not meeting specifications, make sure
that the unit is leak-free by testing all tubing connections
and fit-tings with leak testing solution. Protect circuit board
and all electrical components from leak test solution and
start leak test at the compressor, following air flow to
oxygen outlet. Repair all leaks by tightening connec-tions
and fittings.
7. If unit is alarming, refer to the Alarm Indicator Chart for
probable solutions.
Oxygen Concentrator
23

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents