Service Procedures - CAIRE FreeStyle Comfort Technical Manual

Hide thumbs Also See for FreeStyle Comfort:
Table of Contents

Advertisement

FreeStyle Comfort

Service Procedures

SIEVE BED REPLACEMENT
WARNING: TURN OFF THE UNIT,
DISCONNECT THE POWER CORD
FROM THE ELECTRICAL OUTLET
BEFORE YOU CLEAN THE UNIT TO
PREVENT ACCIDENTAL ELECTRICAL
SHOCK AND BURN HAZARD
REMOVE THE BATTERY FROM THE UNIT BEFORE
PERFORMING ANY MAINTENANCE TO PREVENT
ACCIDENTAL ELECTRICAL SHOCK AND BURN
HAZARD. THE FREESTYLE COMFORT MUST BE
OFF AND DISCONNECTED FROM XTERNAL
POWER BEOFRE REMOVING THE BATTERY PACK
1.
Disconnect the unit from the AC or DC
outlet.
2.
Remove the battery and flip the unit upside
down
3.
Remove the four screws using a T8 torque
bit screwdriver.
4.
Pull the manifold outward and flex the sieve
bed out slightly from the unit so the
manifold can clear the support bolts. Then
simply lift the manifold up and off of the
sieve bed.
24
5.
Flip the unit in the upright position and pull
down on the sieve beds to remove from the
manifold.
6.
Perform the steps in reverse to replace the
old sieve beds with new sieve beds