4. Unplug valve and power cord wire connections from
PCB.
5. Remove power cord by removing the cable tie or pulling
the power cord through the securing cable tie.
6. Carefully remove the tubing from the right side of the O2
monitor of the PCB.
7. Using standard pliers remove the ratchet clamp holding
the braided tubing to the manifold.
Oxygen Concentrator
8. Pull outward on the top of the sieve bed assembly to
remove it from the grommet at the top of the case.
9. Lift upward to remove the beds and product chamber,
while holding camber, from the base of the concentrator.
10. To replace, reverse steps 1-9