Theory Of Operation; Introduction - CAIRE Companion 5 Provider Technical Manual

Oxygen concentrator
Hide thumbs Also See for Companion 5:
Table of Contents

Advertisement

Theory of Operation

Introduction

The Companion 5 is a stationary medical device used to extract
oxygen from the atmosphere, concentrate it, and present the oxygen
to the patient. The device will operate in Continuous Flow Mode. In
Continuous Flow Mode the oxygen is provided at a con-stant flow rate
between 0.5 and 5.0 LPM (Continuously Variable).
Table 2 below lists the major internal components of the concentrator
and their functions. Reference Table 2 and Figure 3 (on the following
page) for clarification while reading the Theory of Operation Section
of the manual.
Item
HEPA Air
Intake Filter
Pump that routes air into and through
Compressor
Cooling Fan
Routes air flow through one sieve bed and
Solenoid
vents the purged air to the atmosphere from the
Valve
other bed. Responsible for alternating flow
Sieve Beds
Routes a larger portion of air exiting one sieve
Orifice
bed back through the other bed for purging.
Routes a smaller portion of air exiting one sieve
Flapper Valve
bed into the product holding tank. Prevents back
flow from product tank to sieve bed
Product
Holds concentrated oxygen prior to its
Holding Tank
regulation and delivery to the patient.
Pressure
Reduces the outlet pressure to be compatible
Regulator
with the flow meter and accessories.
Detects oxygen concentration of flow exiting the
OCSI Sensor
concentrator. Sends a signal to activate alarm if
Flow Control
volumetric flow (LPM) to the patient. Is read
Valve
Oxygen Concentrator
Function
Provides HEPA level filtration for
intake to compressor.
the concen-trator.
Cools the compressor area.
between the beds.
Chemically adsorbs nitrogen
molecules from the air.
concentration is too low.
Integral needle valve that controls
against the printed scale.
HEPA Disc Filter
Responsible for all of the electrical operation of
Printed Circuit
the concentrator. Contains a programmable
Board (PCB)
microprocessor that controls valve timing, alarm
indications, and OCSI functions (if applicable).
Optional Gross
Provides gross particle filtration for intake
Particle Filter
(Top Handle)
Required Gross
Particle Filter
(Bottom)
Table 2: Internal Components of Concentrator
Pressure Swing Adsorption
The CAIRE Companion 5 utilizes the Pressure Swing Adsorption
(PSA) Process to concentrate oxygen gas from ambient air. In the PSA
process, a compressor draws ambient air into the machine through an
intake filter. The compressor then forces the filtered air into a solenoid
valve which directs the air into one of two tanks that are full of a
molecular sieve material, referred to as sieve beds. As the pressure in
sieve bed 1 increases, nitrogen molecules are removed from the
ambient air and are stored in the sieve material. The gas that exits
sieve bed 1 is highly-concentrated oxygen. The majority of this gas
flows through an orifice to sieve bed 2 and is used to purge the stored
nitrogen gas from its sieve material. The remainder of the oxygen gas
is directed through a flapper valve to the product holding tank. Here, it
is stored for delivery to the patient.
Sieve bed 1 continues to pressurize until the sieve material is
completely saturated with nitrogen. At this time the printed circuit
board (PCB) switches the state of the solenoid valve, dumping
pressurized air in sieve bed 1 back into the atmosphere via the purge
muffler. Simultaneously, the valve now directs the compressed air into
sieve bed 2, which has been completely purged of nitrogen gas and is
ready for nitrogen adsorption/oxygen concentration. This cycle
continuously repeats pressurizing and depressurizing the sieve beds,
feeding concentrated oxygen to the product holding tank.
HEPA level filtration of outlet flow.
to compressor.
Provides gross particle filtration for
intake to compressor.
11