Download Print this page
CAIRE Companion 5 Quick Reference Manual

CAIRE Companion 5 Quick Reference Manual

Oxygen therapy
Hide thumbs Also See for Companion 5:

Advertisement

Quick Links

CAIRE Companion 5
5 LPM Compact Stationary with ultraSILENT
Flow Rates
0.5 LPM – 5.0 LPM
Weight
36.0 lb (16.3 kg)
Sound Level
40 dB(A) at 2.0 LPM
Power Consumption
250 Watts at 2.0 LPM,
350 Watts Maximum
ON/OFF
With the unit plugged in, press the power switch upward in the ON (I) position. The audible alarm will sound briefly and all LED lights will be
illuminated for 4 seconds. Then only the Green Light will remain on to show normal operation. If the unit has OCSI (Oxygen Monitoring), its
alarm (yellow light) is disabled for up to 10 minutes while the concentrator warms up.
To turn the unit off, press the power switch downward in the OFF (0) position. All lights will turn off and oxygen flow will cease.
Nasal Cannula and Tubing
Standard oxygen tubing and nasal cannulas can be used with the Companion 5. Maximum tubing length is 50 ft (15.2 m) to be used in
addition to a 7' (2.1m) nasal cannula.
Humidifier Bottle
A humidifier bottle may be used with the Companion 5 at all flow settings. It sits on the stand underneath the outlet barb on the humidifier
stand and is secured with an elastic strap.
Preventative Maintenance
Preventative Maintenance (PM) should be performed every 2 years. It can be performed more frequently as required by the provider based on
their requirements or dirty operating environments. Preventative Maintenance consists of:
• Replace the Air Intake Filter
• Replace the HEPA Filter
Maintenance should not be performed by patients. Note: Step-by-step instructions for these maintenance procedures can be found in the
Provider Technical Manual.
Air Flow and Placement
Ambient air is drawn into the Companion 5 at 2 different locations. The vents on the back draw in air for cooling purposes. Air enters the flow
path underneath the Companion 5 through the intake filter.
Exhaust air exits the Companion 5 through the vents on the lower sides of the case.
Q
R
uick
and autoFLOW
G
efeRence
uide
Technology

Advertisement

loading

Summary of Contents for CAIRE Companion 5

  • Page 1 Air Flow and Placement Ambient air is drawn into the Companion 5 at 2 different locations. The vents on the back draw in air for cooling purposes. Air enters the flow path underneath the Companion 5 through the intake filter.
  • Page 2 HEPA Filter The HEPA filter is located inside of the Companion 5 on the front case half of the unit. It should be replaced once every 2 years or more frequently as needed. Oxygen Concentration The Companion 5 will produce oxygen at 90% (+5.5/-3%) at all flow rates.