Indications for Use

The Companion 5 is indicated for the administration of supple-
mental oxygen. The device is not intended for life support nor does
it provide any patient monitoring capabilities.
A physician must prescribe a specific oxygen flow rate setting to
meet patients' individual needs.
Recommended oxygen flow rates should be adjusted only under the
advice of a physician.
WARNING:
Federal law restricts this device to sell by or on
the order of a physician.

Contraindications

WARNING: The Companion 5 is not intended for life sup-
porting or life sustaining applications, nor does
it provide any patient monitoring capabilities.
WARNING:
In certain circumstances, the use of non-prescribed
oxygen can be hazardous. This device should only
be used when prescribed by a physician.
WARNING:
Not for use in the presence of
flammable anesthetics.
WARNING: As with any electrically powered device, the
user may experience periods of non-operation
as a result of electrical power interruption, or
the need to have the Companion 5 serviced
by a qualified technician.
The Companion 5 is not appropriate for any
patient who would experience adverse health
consequences as a result of such temporary
interruption.
Oxygen Concentrator

Introduction

Welcome to the CAIRE Companion 5 oxygen concentrator. Setting up
and training your patient to use the Companion 5 has never been
easier! You can expect your patients and care providers to easily
learn how to use the device by following the directions in this section.
While setting up and training a patient, be sure to point out the
advantages of the Companion 5. For example:
• Easy-to-use controls
• Quiet operation
• Self-monitoring alarm system
After completing each training procedure, ask your patient if
he or she has any questions. Proper training of your patients
will result in fewer service calls, improved compliance and
increased patient satisfaction.

Pre-Delivery Check List

Verify that the Companion 5 is provided to the patient with the
following items:
• Users' Manual
• Required Oxygen Delivery Accessories (Cannula,
Tubing, Humidifier, etc.)
• Power - Insert the AC power cord into an electrical
outlet to check for proper operation
• Before delivering the device, check and log the status of
the following using the LCD Display:
Hours on the hour meter
Software Revision (The LCD Display will show
this upon startup)
You may adjust the liter flow settings to your patient's
prescription when you deliver and set up the device.
Connecting the AC Power
The Companion 5 operates from external power.
1. Plug the power cord into an AC outlet.
2. The Companion 5 is now ready for use.
If the Companion 5 is not receiving power when
Note:
it is turned on, the unit will alarm and display
a solid red light.
WARNING:
Ensure adequate clearance around the AC Cord.
15