NOYAFA NF-868 Instruction Manual page 22

Multipurpose lcd display cable test & inspection instrument
Hide thumbs Also See for NF-868:
Table of Contents

Advertisement

Unit setup:
Set unit: meter
When moving cursor
is shown. Move the cursor
to the main menu. The following interface will be shown:
Note:
Setup of the unit of inch and yard is just the same as that of meter.
Calibration function:
Due to different materials of the cables, calibration is required before
testing the cables. The function of calibration is for accurately measuring
length of the cables. It is a must to perform calibration as stated below and
the cable length used for calibration is required to be over 10 meters.
When starting dynamic calibration, insert the same type of cables at a
specified length into "M" port. It is unnecessary to insert into the far end.
Select the main menu. When selecting cable type and unit ready, move
cursor
"
"
to "Calibration parameter" item, and then press "ENTER" key.
The following interface will be shown:
Select "Yes", and then press "ENTER" key, the measured length will be shown:
At this time, press "UP" or "DOWN" key to adjust to show the actual length
as below:
When length is given, press "ENTER" key to save the calibration value. And
then exist from the calibration function, and return to the main menu.
"
"
to "Unit" item, press "ENTER" key till unit of meter
"
"
to "Return" item, press "ENTER" key to return
Type: PAIR(8)
U n i t : M e t e r
C a l i b r a t i o n
R e t u r n
C a l i b r a t i o n
No
- -
B a s e A d j u s t
10.5m
-19-