Advertisement

Quick Start Guide
2

Installation

2.1
General installation information
WARNING
Electrical shock
Installation and servicing of this product may expose personnel to
dangerous voltages.
Electrical installation must be in accordance with the National Electrical
Code (ANSI/National Fire Protection Agency-70) and/or any other
national or local codes.
The equipment is protected throughout by double insulation.
Disconnect main power, wired to separate power source, before
servicing.
Ensure that all equipment doors are closed and protective covers are in
place, except when maintenance is being performed by qualified
personnel.
Signal wiring must be rated at least 240 V.
Non-metallic cable strain reliefs do not provide grounding between
conduit connections. Use grounding type bushings and jumper wires.
Make sure that the instrument is connected and properly grounded
through a three-wire power source.
Proper use and configuration is the operator's responsibility.
1. Install the transmitter with a sun shield or out of direct sunlight and
areas with extreme temperatures.
2. Install the system in an area where vibrations and electromagnetic
and radio frequency interference are minimized or absent.
3. Keep the transmitter and sensor wiring at least 1 ft (0.30 m) from
high voltage conductors. Be sure there is easy access to the
transmitter and sample conditioning system.
4. The transmitter is suitable for panel, pipe, or surface mounting.