Sartorius WZA N Series Installation Instructions Manual page 28

Weigh cells
Hide thumbs Also See for WZA N Series:
Table of Contents

Advertisement

Special Codes
Position 1
2
3
4
5
*
*
*
*
*
or
or
or
*:
Space
– –:
Final readout
H:
Overload
H H:
O verload in checkweighing
(Function is only available during operation with following peripheral devices:
Optional display unit or software YAD01IS)
L:
Underweight
L L:
Underweight in checkweighing
C:
Adjustment
Error Messages
Position 1
2
3
4
5
*
*
*
E
r
*:
Space
# # #:
Error number
Example: Output of the weight value + 1255.7 g
Position 1
2
3
4
5
+
*
*
*
1
Position 1:
Plus +, minus –, or space
Position 2:
Space
Position 3 - 10:
Weight value with decimal point; leading zeros are printed as spaces
Position 11:
Space
Position 12 - 14:
Characters for unit of measure or space
Position 15:
Carriage return
Position 16:
Line feed
Data Input Format
A computer connected via the data port can send commands to the device to control device
functions.
The commands sent are control commands and may have different formats.
Control commands have up to 26 characters. Each of these characters must be
sent based on the setup configuration for data transmission.
Formats for Control Commands
Format 1:
Esc
!
CR
Format 2:
Esc
!
#
Esc:
Escape
#:
Number
_:
Underscore (ASCII: 95)
CR:
Carriage return (optional)
LF:
Line feed (optional)
max:
Depending on the command character, i.e. parameter: The entry is truncated after
the max. length, and not rejected as when entered via the keyboard
28
6
7
8
9 10 11 12 13 14 15 16
*
*
*
*
*
H H
L
L
C
6
7
8
9 10 11 12 13 14 15 16
r
*
# # #
*
*
6
7
8
9 10 11 12 13 14 15 16
2
5
5
.
7
*
g
LF
_
CR
LF
!:
Command character
*
* CR LF
*
* CR LF
*
* CR LF
Format 1 (e.g., ESC K)
!
Meaning
K
Filter adjustment:
Very stable conditions
L
Filter adjustment: Stable conditions
M
Filter adjustment:
Unstable conditions
N
Filter adjustment:
Very unstable conditions
O
Lock keys
Q
Acoustic signal (beep)
P
Print
R
Release keys
S
Restart
T
Tare and zero
Z
Internal adjustment
Format 2 (e.g., ESC f3_)
!#
Meaning
f1_ Calibrate or Adjust according to
menu setting
f3_ Zero
f4_ Tare (without zeroing)
s1_ External adjustment
s3_ Function [CF]
x0_ Perform internal calibration
x1_ Print load cell type
x2_ Print load cell series no.
x3_ Load cell software version