10.1.3 Tubes and hoses of HVE and saliva ejector
CAUTION
Tubes and hoses of HVE and saliva ejector should be cleaned after treating each
patient.
▶ Otherwise, odor may occur and particles in hoses may coagulate which may
1) Turn on the main switch.
2) Aspirate 1~2 cups of water through HVE and saliva ejector respectively.
3) Put back instruments to the holder.
Wash tubes and lines in accordance with the following procedure by using a chemical
cleaning agent, Orotol plus, recommended by Shinghung.
1) Aspirate 2 liter of water before disinfection.
2) Mix 2 liters of a 2% strength solution of Orotol_plus.
Fill the OroCup up to the 2 liter mark with cold water,
add 40mL Orotol_plus and mix thoroughly.
3) Attach suction hoses to the care system and suck up
one liter of the solution per unit.
4) Allow to react for at least 1 hour or overnight.
5) Disinfect the spittoon with the remainder of the solution.
6) After the reaction time/next morning, suck up 2 liter
of water.
Revised on 15/06/2021
weaken the suction power.