Doc -030
8.7 Volume adjustment of air/water on instruments
8.7.1 Instruments in dentist table
▶ Adjust volume of each instrument with a control knob located in the front
lower of the table
AIR
①
8.7.2 3-WAY syringe in assistant table
▶
To adjust the air/water volume for a 3-WAY
syringe in the assistant table, open the
panel in the left of the table and
adjust the valve mounted on the right side.
8.7.3 HP oil collector
▶ Located at the bottom of the dentist table.
▶ The container collects water and lubrication oil
particles generated from high speed handpieces.
Check inside monthly and dispose the debris collected
Revised on 15/06/2021
.
3-WAY
TURBINE
WATER
②
③
Knob No
①
Air volume for holder 1
②
Water volume for holder 1
③
Water volume for holder 2
④
Water volume for holder 3
⑤
Air volume for holder 4
⑥
Water volume for holder 4
⑦
Water volume for holder 5
Hatch comfort C1 User manual
MOTOR
AIR
WATER
④
⑤
⑥
Function
1) The knob location is
2) Turn the knob
..
SCALER
⑦
Remark
based on the standard
layout of instruments.
The position of a knob
may vary based on a
layout change.
counterclockwise to
increase the volume.
Air adjustment valve
Water adjustment valve
Oil Collector