Advertisement

Quick Links

Doc -030
Hatch comfort C1 User manual
User Operating Manual
Revised on 15/06/2021

Advertisement

Table of Contents
loading

Summary of Contents for hatch comfort TAURAUS C1

  • Page 1 Doc -030 Hatch comfort C1 User manual User Operating Manual Revised on 15/06/2021...
  • Page 2: Table Of Contents

    Doc -030 Hatch comfort C1 User manual Table of contents Description ---------------------------------------------------------------------------------------------- User Instructions ---------------------------------------------------------------------------- Transport and Storage ---------------------------------------------------------------------- Safety ----------------------------------------------------------------------------------------------------- Safety Indication ------------------------------------------------------------------------------- Right use of way to operate unit ------------------------------------------------------------- Safety Guide ---------------------------------------------------------------------------------- Installation --------------------------------------------------------------------------------------------- Connecting of Instruments...
  • Page 3: Description

    Doc -030 Hatch comfort C1 User manual 1. Description TAURAUS C1 is a dental chair and units to be used to take for diagnosis and therapy of patients. It consists of the main unit and accessories. The dental chair is used for patient to sit during the dental diagnosis, treatment and /or operation.
  • Page 4: User Instructions

    Doc -030 Hatch comfort C1 User manual 1.1 User instructions Recommendation It is recommended to read the manual thoroughly before the operation of product in order to prevent damage or wrong use. 1.1.1 Summary Short term Full description Technical instructions...
  • Page 5: Transport And Storage

    Doc -030 Hatch comfort C1 User manual 1.2 Transport and storage NOTE When performing repair and service work, keep transport and storage condition. Below labeled points should be attached on the surface of packing boxes and should be observed. Stand upside in the direction of arrow mark Fragile.
  • Page 6: Safety

    Doc -030 Hatch comfort C1 User manual 2. Safety 2.1 Safety Indication This manual shows symbols as below in the contents to enable operators to use products in the safe and proper way to prevent any property damage or physical injury in advance.
  • Page 7: Right Use Of Way To Operate Unit

    “Proper way to operate” refers to the instruction for use, inspection and the scope of warranty service. 1) The responsibility of Hatch Group Inc. for the product performance, reliability and safety: Hatch Group Inc. manufacturers its products in compliance with domestic and international standard.
  • Page 8: Safety Guide

    If the dental chair unit has been out of operation for more than a few days, it should be inspected and cleaned according to the instructions. Use only genuine parts provided by Hatch Group Inc.. 2.3 Safety Guide 2.3.1 General information...
  • Page 9 Doc -030 Hatch comfort C1 User manual 2.3.2 Features of product CAUTION Avoid an injury to the patient or operator while the device is in use. ▶ Pay attention when the operating unit is in use. ▶ Do not stand on the chair and jump down.
  • Page 10 Doc -030 Hatch comfort C1 User manual CAUTION Do not apply an excessive load to the tumbler filler. Do not operate or disassemble the unit with an excessive force. WARNING ▶ Do not touch the light head lamp or adjacent parts during and after the operation.
  • Page 11: Installation

    NOTE The installation of this dental chair unit should be done in accordance with the installation manual by Hatch Group Inc. technicians or authorized dealers only. WARNIG ▶ In order to avoid electrical shock, a protective earth should be also connected when connecting the power to this dental chair unit.
  • Page 12: Connecting Of Instruments

    Doc -030 Hatch comfort C1 User manual 4. Connecting of instruments The dental chair unit can be connected with handpiece instruments. Types of instruments can connect to the dental chair unit are shown in 8.6.5. CAUTION All instruments should be connected to tubing before use.
  • Page 13: Dental Chair Unit

    Doc -030 Hatch comfort C1 User manual 6. Dental chair unit operation Do not put any objects underneath seat or place hands or other body parts to moving part such as the backrest. ▶ Doing so, it may cause physical injury or property damage/ breakdown of...
  • Page 14: Product Composition

    Doc -030 Hatch comfort C1 User manual 7. Product composition 7.1 Main composition Components Components Foot controller Tumbler filler Backrest Patient chair Headrest Chair base Assistant table Seat Spittoon Dentist table Operating lamp Table arm Operating lamp arm Unit ▶ Applied part : Upholstery in headrest, backrest and seat, Instrument tips...
  • Page 15: Packaging

    Doc -030 Hatch comfort C1 User manual 7.2 Packaging Carton 1 Carton 2 Carton 3 Operating lamp & Arm Chair Stool foam Table arm Foot controller Dentist table Armrest Backrest Unit Seat Assistant table Headrest Junction box Spittoon & Manual...
  • Page 16: Functions

    Doc -030 Hatch comfort C1 User manual 8. Functions 8.1 Turning on the main power 8.1.1 Connecting the main power plug ▶ Confirm the operating voltage of the dental chair unit. ▶ Connect the power plug to a power outlet.
  • Page 17: Junction Box

    Doc -030 Hatch comfort C1 User manual 8.2 Junction box CAUTION Do not put objects or step on the junction box. A cover may be broken and cause property damage or physical injury. 8.2.1 Function of junction box ▶ The station connects water and air from the floor to the dental chair unit while shut-off valves are installed on top of it ①...
  • Page 18: Chair

    Doc -030 Hatch comfort C1 User manual 8.3 Chair CAUTION Injury may occur during the movement of the chair to a programmed position. ▶ Only an authorized person shall operate the dental chair unit. NOTE A thermo-sensor is installed in hydraulic pump for seat and backrest. If the chair is in use continuously for a long period of time, the chair operation may stop to cool down.
  • Page 19 Doc -030 Hatch comfort C1 User manual 8.3.1 Operation of movable armrest (Option) ▶ This section is applicable to the movable armrest.(right side of the chair) ▶ The armrest can be swiveled by 160 degree for the patient’s convenience to enter or exit from the chair.
  • Page 20 Doc -030 Hatch comfort C1 User manual 8.3.2 Operation of headrest ▶ The height and angle of the headrest should be adjusted according to the patient’s size and the treatment position. CAUTION Do not overload the headrest and adjacent parts.
  • Page 21 Doc -030 Hatch comfort C1 User manual CAUTION ▶ If the knob is not tightened firmly, the headrest may move during an operation and cause an accident. ▶ When turning the knob, pay attention to prevent fingers or hair from being caught in the knob.
  • Page 22: Dentist Table

    Doc -030 Hatch comfort C1 User manual When the backrest moves down When the seat moves down 8.4 Dentist table NOTE The instrument layout can be rearranged depending on the operator’s preference and may be different from the picture below.
  • Page 23 Doc -030 Hatch comfort C1 User manual CAUTION Do not overload the table more than 30N(3㎏∙f). Do not operate with excessive force and avoid a shock. ▶ Overload may cause the table to go down resulting injury or breakdown. NOTE The tray’s silicone pad should be sterilized regularly for 3 minutes under 134℃...
  • Page 24: Assistant Table

    Use only the syringe tip provided by HATCH GROUP INC. which have been engineered and manufactured for HATCH GROUP INC. syringes. Using other syringe tips other than the one provided by HATCH GROUP INC. may result in the syringe tip ejection.
  • Page 25: Control Panel

    Doc -030 Hatch comfort C1 User manual Description Description ① Control panel ④ Saliva ejector ② ⑤ 3-WAY syringe Assistant tray ⑥ Instruments holder ③ CAUTION Do not sit on the table or move it with excessive force. CAUTION Except 3-WAY syringe, all other instruments should be placed in the holder after use.
  • Page 26 Doc -030 Hatch comfort C1 User manual ① [ Dentist panel (Option) ] [ Dentist panel (standard) ] A : Chair position control ① : programmed positions ② : manual positions B : Function and menu control C : LCD screen [ Assistant panel ] 8.6.2 Chair control...
  • Page 27 Doc -030 Hatch comfort C1 User manual Zero position Chair moves to the patient entry/exit position. Chair moves to Auto Position 1 programmed AP 1 for treatment. Chair moves to Auto Position 2 programmed AP 2 for treatment. Chair moves to the previous position.
  • Page 28 Doc -030 Hatch comfort C1 User manual ▶ Function of each switch function marked * works only when instruments are picked up Symbol Description Function Display (option) Operating Turning on/off the operating light light control Tumbler water Water is supplied to tumbler while supply pushing this switch.
  • Page 29 ▶ otherwise, an accident or breakdown may occur. CAUTION Use only handpieces recommended by Hatch Group Inc.. The connection between the handpiece tubing joint and coupling should be Mid West Type 4-Hole, Borden 2-Hole or 3-Hole.
  • Page 30 Doc -030 Hatch comfort C1 User manual 8.6.5 Instruments recommended by Hatch Group Inc. Category Models Connector type Brand Remark SUCTION SYRINGE Suction handpiece large Standard Durr (HVE) SALIVA EJECTOR Suction handpiece small Standard Durr SYRINGE SCALER Piezon kit Standard NOTE ▶...
  • Page 31 Doc -030 Hatch comfort C1 User manual 4) Press the timer switch to switch on / off each holder position. Press switch to select the holder position from 2 through 5. By pressing the timer switch again, switch on I (holder ON). Pressing the timer again, switch off O (holder OFF).
  • Page 32: Volume Adjustment Of Air/Water On Instruments

    Doc -030 Hatch comfort C1 User manual 8.7 Volume adjustment of air/water on instruments 8.7.1 Instruments in dentist table ▶ Adjust volume of each instrument with a control knob located in the front lower of the table 3-WAY TURBINE MOTOR...
  • Page 33: Spittoon Flushing & Tumbler Filler System

    Doc -030 Hatch comfort C1 User manual 8.8 Spittoon rinsing & tumbler filler system 8.8.1 Spittoon bowl rinsing CAUTION Don’t let the patient hold on to the spittoon bowl, rinsing pipe or tumbler filler in order to prevent damage or injury.
  • Page 34 Doc -030 Hatch comfort C1 User manual 8.8.2 Tumbler filling CAUTION Only a cup made of paper, metal or opaque plastic can be used. ▶ The tumbler filler sensor does not work with transparent glass or plastic cups. WARNING If water in the tumbler overflows, water may ingress into the dental chair unit which may cause malfunction or electrical shock.
  • Page 35 Doc -030 Hatch comfort C1 User manual 8.8.3 Water heater (Warmer) WARNING If the water heater operates without water supply, it may be overheated, damaged and cause fire. When the heater is not in use, turn OFF the heater switch.
  • Page 36: Foot Controller

    Doc -030 Hatch comfort C1 User manual 8.9 Foot controller ① ② ⑤ ③ ④ ▶ Using the foot controller levers, control the chair seat and backrest. Function How to ① Backrest DOWN While pushing out, the backrest moves down.
  • Page 37: Operating Light

    Doc -030 Hatch comfort C1 User manual 8.10 Operating light 8.10.1 The structure of operating light WARNING ▶ Do not apply excessive shock or vibration to the operating light (head, arm). Prevent it from getting wet by moisture. If wet, turn off the main switch and wipe the moisture with a dry towel.
  • Page 38 Doc -030 Hatch comfort C1 User manual 8.10.2 Operation NOTE The operating light should be turned off when not in use. ▶ The factory setting let the operating light turns ON automatically when a handpiece (except 3-way syringe) is picked up from the instrument holder.
  • Page 39 Doc -030 Hatch comfort C1 User manual 8.10.3 Focus adjustment CAUTION The range of the adjustment cap movement is limited. If the cap movement gets stiff, stop turning it. If turned forcibly, it may be damaged and dismantled. ▶ The operating light is configured at the standard distance of 700mm from the patient mouth.
  • Page 40 Doc -030 Hatch comfort C1 User manual Light intensity depending on filament direction Category Vertical Horizontal Low mode about 7,500 ㏓ Light intensity about 24,000 ㏓ High mode 130㎜ 170㎜ Illumination range 70㎜ 45㎜ Remark) The light intensity shown is based on the distance of 700mm from the light source to an object.
  • Page 41 Doc -030 Hatch comfort C1 User manual 8.10.6 Adjusting the operating light head ▶ To adjust the angle of the illumination easily, all arm joints can be rotated in 3 dimensions. When adjusting the position and direction, hold the light handles with both hands and move it gently.
  • Page 42 Doc -030 Hatch comfort C1 User manual 8.10.7 Replacement of lamp (bulb) ▶ When replacing the halogen lamp, always use genuine parts. NOTE Specifications of the halogen lamp : AC 12V, 50W. socket type ( E11 ) NOTE Oil from fingers may shorten the lamp lifespan.
  • Page 43: Distilled Water Tank System (Option)

    Doc -030 Hatch comfort C1 User manual 8.10.8 Cleaning and disinfection CAUTION The operating light is hot after use. Clean and disinfect the lamp after it cools down completely. CAUTION A special care should be taken for the reflector and light cover to avoid scratches during cleaning and handling.
  • Page 44 Doc -030 Hatch comfort C1 User manual CAUTION When distilled water will not be used over night or for a few days, turn ‘OFF’ the air valve. ▶ If the valve remains ‘ON’ over night or a few days, the air pressure in the bottle becomes excessive and may damage the parts.
  • Page 45: Spittoon Bowl Valve (Option)

    Doc -030 Hatch comfort C1 User manual 8.12 Spittoon bowl valve (Option) ▶ The spittoon bowl valve contains and transports what drained out from the spittoon bowl to the amalgam separator integrated in the suction motor. It automatically drains out if certain amount of liquid is collected inside of the valve and also prevent the spittoon bowl from being connected directly to the suction line where negative pressure exists so that noise and odor does not come out of the spittoon bowl.
  • Page 46: Cart-Type Table (Option)

    Doc -030 Hatch comfort C1 User manual 8.13 CART-TYPE table (option) 8.13.1 Structure ※ chair and unit part is same as Over-arm type. Description Description Dentist table Caster Column Height adjustment knob Base Flexible conduit Caster (with brake) 8.13.2 Function 8.13.2.1 Caster breaker...
  • Page 47 Doc -030 Hatch comfort C1 User manual 8.13.2.2 How to adjust the table height ▶ With one hand, push down the top center of the dentist table. With the other hand, turn the height adjustment knob counterclockwise. ▶ To raise the table, release the hand from the table. The table goes up and tighten the knob clockwise at the height you want to fix.
  • Page 48: Fuse Replacement

    Doc -030 Hatch comfort C1 User manual 9. Fuse replacement CAUTION If a fuse in the dental chair unit is burnt out, contact your local sales distributors. WARNIG Only a certified or authorized person can perform repair and technical services.
  • Page 49: Fuse For Main Transformer

    Doc -030 Hatch comfort C1 User manual 9.2 Fuse for main transformer Fuse (4A / 250V) A fuse is located inside the chair base. The fuse sizes are Φ6㎜ diameter, 30㎜ length. Main transformer fuses should be replaced in the following sequence.
  • Page 50: Maintenance

    Doc -030 Hatch comfort C1 User manual 10. Maintenance WARNING When performing cleaning or technical service work, turn OFF the main switch. Also close water and air shut-off valves, if necessary. ▶ If the main switch is not OFF, electrical shock may occur and cause a serious injury or breakdown.
  • Page 51 Doc -030 Hatch comfort C1 User manual 10.1.2 Disinfection with alcohol CAUTION Use alcohol for disinfection of joints, axis and switches. ▶ Other types of cleaning solution may cause discoloration or damage to the chair materials. Do not let alcohol smear into the dental chair unit.
  • Page 52 Doc -030 Hatch comfort C1 User manual 10.1.3 Tubes and hoses of HVE and saliva ejector CAUTION Tubes and hoses of HVE and saliva ejector should be cleaned after treating each patient. ▶ Otherwise, odor may occur and particles in hoses may coagulate which may weaken the suction power.
  • Page 53 Doc -030 Hatch comfort C1 User manual 10.1.4 Suction filter CAUTION Clean the suction filter at least once daily . CAUTION When cleaning the assembling filter cap, ensure that the rubber ring does not come out and snap. The cap should be tightly contacted.
  • Page 54 Doc -030 Hatch comfort C1 User manual DANGER There may be infectious waste. Follow infection control regulations to prevent cross contamination. 10.1.5 Spittoon bowl and spittoon trap CAUTION Clean the spittoon trap at least once daily. The spittoon bowl should be cleaned as needed.
  • Page 55 Doc -030 Hatch comfort C1 User manual Rinsing pipe ▶ Disassembling rinsing pipe knob When disassembling the rinsing pipe, Rinsing pipe Loosen the knob and pull out the pipe. Assemble in the reverse order. CAUTION When disassembling the rinsing pipe, pay attention not to damage or lose the rubber ring.
  • Page 56 Doc -030 Hatch comfort C1 User manual ▶ 3-WAY syringe Pick up syringes from both the dentist and assistant holder and press levers to eject water into the spittoon bowl. ▶ Air turbine and micromotor 1) Pick up a handpiece from the holder.
  • Page 57 Doc -030 Hatch comfort C1 User manual Do not apply a floor-cleaning detergent or wax on the base of the dental chair unit and the foot controller. If stained, wipe if off immediately. ▶ These materials may cause damage on parts by discoloration and breakdown.
  • Page 58: Monthly

    Doc -030 Hatch comfort C1 User manual ▶ Other parts and surface CAUTION Clean frequently as necessary. 1) Wipe off with a sponge or soft fabric soaked with a neutral detergent. 2) Wipe off moisture and detergent on the surface completely with dry fabrics.
  • Page 59 Doc -030 Hatch comfort C1 User manual 10.2.2 HP lubrication oil particle collection bottle CAUTION Check the condition of the bottle located in the bottom of the dentist table at least monthly. CAUTION Oil absorbed by a filter is flammable. Dispose the filter according to the local regulation.
  • Page 60: Every 6 Month

    Doc -030 Hatch comfort C1 User manual 10.3 Every 6 month 10.3.1 Water and Air filter NOTE. The water and air filter should be cleaned at least once in every 6 month. Clean as necessary, if the water pressure gage shows below 0.17MPa or the air pressure gage shows 0.39MPa or the volume of air/water supplied to handpiece...
  • Page 61: Care And Asepsis Instructions

    Transfer surfaces are the areas that are contaminated by instruments and other inanimate objects. HATCH GROUP INC. recommends a barrier protection for all applicable touch and transfer surfaces. When used, barriers must be FDA market-cleared barrier plastic. DANGER Cover-All barrier film or any other FDA market-cleared barrier film is suitable for this application.
  • Page 62: Before Claiming Services

    Doc -030 Hatch comfort C1 User manual 11. Before claiming service When problems occur, please refer to troubleshooting as shown below. If problems persist or observe problems other than described herein, stop the operation, close the AIR / WATER shut-off valves and contact a local distributor.
  • Page 63: Electromagnetic Compatibility(Emc)

    Doc -030 Hatch comfort C1 User manual 12. Electromagnetic Compatibility (EMC) Medical electrical dental chair unit needs special precautions regarding EMC and needs to be installed and Put into service according to the EMC information provided in this manual. Portable and mobile RF communications dental chair unit can affect medical electrical dental chair unit.
  • Page 64 Doc -030 Hatch comfort C1 User manual Immunity IEC 60601 Electromagnetic environment – Compliance level test test level guidance Mains power quality should be that of a <5% U <5% U typical commercial or hospital environment. (>95%dip in U (>95%dip in U...
  • Page 65 Doc -030 Hatch comfort C1 User manual NOTE 1 is the a.c mains voltage prior to applications of the test level. The TAURUS C1 is intended for use in the electromagnetic environment specified below. The customer or the user of the TAURUS C1 should assure that it is used in such an environment.
  • Page 66: Specification

    Doc -030 Hatch comfort C1 User manual 13. Specification 1) Specification Rated voltage and frequency AC 100/110V/120V, AC 210V/220V/230V, 50/60HZ Power consumption 1,000VA Type of protection against electrical shock Class I Degree of protection against electrical shock Grounding earth 3 phase earth Input water pressure 0.2 ~ 0.59MPa...
  • Page 67: Warranty

    Doc -030 Hatch comfort C1 User manual Manufacturer: Hatch Group Inc. 4181 Sladeview Crescent Unit 36 Mississauga ON L5L5R2 Canada http: //www.hatchgroup.ca Revised on 15/06/2021...
  • Page 68 Doc -030 Hatch comfort C1 User manual Revised on 15/06/2021...

Table of Contents