Chapter 6, Maintenance, of the manual for more detailed information regarding the zeroing
procedure of the ADTS.
The ADTS can be zeroed at higher vacuum levels, however, the uncertainty of vacuum sensors
tend to increase at higher vacuum levels, therefore, the uncertainty in the ADTS would also
increase if zeroed at higher vacuum levels.

Menu | Test

The Menu | Test section of the system is used to perform a number of the diagnostic and tuning
functions that are available with the RUSKA 7750i. The Sweep and Control test operate on the
current selected channel. The other screens are independent of the selected channel.
Menu | Test — Self Test
The ADTS can perform electrical and pneumatic self-test to assist in trouble shooting potential
problems. Please refer to Chapter 6, Maintenance, for more details.
Menu | Test — Remote Test
The Menu | Menu | Test | Remote section of the system is used to perform a number of the
diagnostic functions on the remote interface. It can be used to display the transmitted and
received messages that are sent across the interface. This can be a powerful tool to assist in
identifying the source of communication problems.
The Menu | Menu| Test | Remote menu displays information from the IEEE-488 interface.
Pressing the Serial 1 [F2] or Serial 2 [F3] keys will display information from the serial ports.