LINET AVE 2 User Manual And Technical Description page 26

Table of Contents

Advertisement

portable and mobile RF communications equipment and the birthing bed AVE2
The birthing bed AVE 2 is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the AVE 2 bed can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
birthing bed AVE 2 as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum
output of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Recommended separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
0,12
0,37
1,1
3,6
11
m
80 MHz to 800 MHz
0,40
1,2
4
12
40
800 MHz to 2,5 GHz
0,40
1,2
4
12
40
26