Instructions for Use and
Air2Care IN & HybriMatt
Integrated Mattress System for Essenza 300
with Air2Care 6 IN, Air2Care 8 IN air mattresses and
HybriMatt
®
Technical Description
100 IN hybrid foam mattress
100 IN
®
D9U00A2CI-0101
Version: 02
Publication Date: 2025-01

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Air2Care IN and is the answer not in the manual?

Questions and answers

Summary of Contents for LINET Air2Care IN

  • Page 1 Instructions for Use and Technical Description Air2Care IN & HybriMatt 100 IN ® Integrated Mattress System for Essenza 300 with Air2Care 6 IN, Air2Care 8 IN air mattresses and HybriMatt 100 IN hybrid foam mattress ® D9U00A2CI-0101 Version: 02 Publication Date: 2025-01...
  • Page 2 274 01 Slaný Tel.: +420 312 576 111 Fax: +420 312 522 668 E-mail: info@linet.cz http://www.linet.com Service department: service@linetgroup.com Air2Care IN & HybriMatt 100 IN Integrated Mattress System for Essenza 300 Author: LINET, s.r.o. Related links: www.linet.com D9U00A2CI-0101 Version: 02 Publication Date: 2025-01 Copyright ©...
  • Page 3: Table Of Contents

    11.7 Handling and Storage..............................42 11.8 Correct Use of the Air2Care IN mattresses with Air2Care IN & HybriMatt 100 IN compressor ........43 11.9 Correct Use of the HybriMatt 100 IN mattress with or without Air2Care IN & HybriMatt 100 IN compressor....46 D9U00A2CI-0101_02...
  • Page 4 17.1 Mechanical Specifications of the Air2Care IN Mattresses ..................59 17.2 Mechanical Specifications of the HybriMatt 100 IN Mattress ..................60 17.3 Mechanical Specifications of the Air2Care IN & HybriMatt 100 IN Compressor ............60 17.4 Electrical Specifications of the Air2Care IN & HybriMatt 100 IN Compressor ............. 61 17.5 Air2Care IN Environment Conditions ..........................
  • Page 5 Fig. SCU side view (from the foot end of the bed towards the head end when the SCU is integrated to the bed) . 26 Fig. Installation of the 3 Air Hoses to the Air2Care IN & HybriMatt 100 IN compressor ..........27 Fig.
  • Page 6: Symbols And Definitions

    1 Symbols and Definitions 1.1 Warning Notices 1.1.1 Types of Warning Notices Warning notices are differentiated by the type of danger using the following signal words: ► CAUTION warns about the risk of material damage. ► WARNING warns about the risk of physical injury. ► DANGER warns about the risk of fatal injury. 1.1.2 Structure of Warning Notices WARNING! Type and source of risk! ►...
  • Page 7: Abbreviations

    1.4 Abbreviations AC ( ~ ) Alternating Current Alternating Pressure Therapy (Mode) European Conformity Cardiopulmonary Resuscitation Sound Intensity Unit Direct Current DC ( Constant Low Pressure (Mode) Configuration number Electromagnetic Compatibility Field-effect transistor High Frequency High Pressure Laminate Intensive Care Unit INT.
  • Page 8: Important Safeguards And Statements

    2 IMPORTANT SAFEGUARDS AND STATEMENTS When using electrical products, especially when children are present, basic safety present, basic safety precautions should always be followed, including the following: WARNING! If any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority of the Member State in which the user and/or patient is established! WARNING! To reduce the risk of electrocution:...
  • Page 9 HybriMatt 100 IN mattresses is allowed. WARNING! An additional multiple socket-outlet or extension cord shall not be connected to the medical electrical system of the Essenza 300 bed with Air2Care IN & HybriMatt 100 IN compressor. WARNING! Risk of injury due to incorrect use! ►  ...
  • Page 10: Introduction

    3 INTRODUCTION This manual should be used for the initial set up of the system and for reference purposes. 3.1 Product Label and Technical Label The legal label and UDI label are located on the back of the SCU (system control unit). The serial number, model number, and the input rating can be found on the label.
  • Page 11: Fire Resistance Label

    3.1.2 Fire Resistance Label Fig. Fire Resistance Label for Air2Care 6 IN Fig. Fire Resistance Label for HybriMatt 100 IN mattress and Air2Care 8 IN mattress mattress D9U00A2CI-0101_02...
  • Page 12: Compatibility Of Mattresses And Compressors

    Matt 100 IN compressor. Air2Care 6 IN mattress, Air2Care 8 IN mattress and HybriMatt 100 IN mattress are compatible with the same Air2Care IN & HybriMatt 100 IN compressor but some functions are available only for the Air2Care 6 IN mattress or Air2Care 8 IN mattre- Exclusive functions of the Air2Care 6 IN mattress and Air2Care 8 IN mattress: ■   Deflating an individual cell...
  • Page 13: Intended Use Of The Air2Care In Integrated Mattress System

    MATTRESS SYSTEM The intended use of the Air2Care IN integrated mattress system is to provide a support surface for patients by pressure redistribution designed for management of tissue loads. The Air2Care IN mattresses are intended to be used with Air2Ca- re IN &...
  • Page 14: Intended Use Of The Hybrimatt 100 In Integrated Mattress System

    This mattress is intended to be used with Air2Care IN & HybriMatt 100 IN integrated air compressor in the Essenza 300 medical bed. It can improve resting and com- fort of the patient.
  • Page 15: Product Description

    7 PRODUCT DESCRIPTION 7.1 Air2Care IN Mattress (Type B Applied Part) The Air2Care IN & HybriMatt 100 IN compressor comes with cell type mattress of different dimensions and constructions, provided optionally with the SCU. Please refer to the specifications section for details.
  • Page 16: Air2Care 8 In Mattress

    7.1.2 Air2Care 8 IN Mattress 8-inch replacement system with cell-on-cell structure. The 20 transverse air cells in the air layer alternate in pressure, with the exception of 3 static head cells. CPR valves locate at the head section in case rapid deflation is required. Loops are implemented at the base of the mattress for air cells fixation.
  • Page 17: Set Of Mattress Cells

    7.1.3 Set of mattress cells Each Air2Care IN mattress consists of 3 static mattress cells at mattress head end and 17 alternating mattress cells. TRUNK ZONE TRUNK ZONE FOOT FOOT HEAD HEAD ZONE ZONE ZONE ZONE Fig. Structure of the 20 mattress cells 1.
  • Page 18: Fig. Disconnecting A Hose From Mattress Cell

    Fig. Disconnecting a hose from mattress cell Fig. Description of connecting elements 1. Quick release element of the mattress air hose system 2. Fastening element connecting a mattress cell to the mattress base cover To deflate an individual alternating mattress cell in mattress foot zone: ►...
  • Page 19: Hybrimatt 100 In Mattress For Air2Care In & Hybrimatt 100 In Compressor

    In the Passive Mode, HybriMatt 100 IN mattress is used as a foam mattress without air support depending on the integra- ted compressor. To change the HybriMatt 100 IN mattress from active mode to passive mode: ► Unlock the Air2Care IN & HybriMatt 100 IN Control Panel on the Essenza 300 foot plastic siderail by Lock Button. ► Turn off the integrated compressor by the ON/OFF button on the Air2Care IN & HybriMatt 100 IN Control Panel. ►...
  • Page 20: Description Of The Hybrimatt 100 In Mattress

    5. CPR Valve position 6. Mattress Foam Frame 7. Air Tubing 8. Base Mattress Cover Fig. HybriMatt 100 IN mattress for Air2Care IN & HybriMatt 100 IN 9. Three Air Hoses with Connectors to the compressor Air2Care IN & HybriMatt 100 IN compressor...
  • Page 21: Fig. Description Of Hybrimatt 100 In Mattress Bottom Side

    2. Safety Straps with side release buckles (to fix mattress on the bed mattress support platform) 3. Transport Handles 4. Three Air Hoses with connectors to the inte- grated Air2Care IN & HybriMatt 100 IN compre- ssor Fig. Description of HybriMatt 100 IN mattress bottom side D9U00A2CI-0101_02...
  • Page 22: Fig. Mattress Viscoelastic Foam Layer

    Fig. Mattress Viscoelastic Foam Layer Fig. Mattress Structure with highlighted Cell Sleeve with eight cells D9U00A2CI-0101_02...
  • Page 23: Fixation Of The Hybrimatt 100 In Mattress On The Essenza 300 Mattress Support Platform

    7.2.3 Fixation of the HybriMatt 100 IN mattress on the Essenza 300 mattress support platform HybriMatt 100 IN mattress is designed for Essenza 300 mattress support platform. HybriMatt 100 IN mattress is equipped with straps intended for fixing mattress on the Essenza 300 mattress support platform. Fig.
  • Page 24 To fix the  safety straps with side release buckles : ► Lock the side release buckle by connecting its male and female part together. To release the safety straps with side release buckles : ► Release the buckle by pressing it from both sides and by disconnecting its male and female part. D9U00A2CI-0101_02...
  • Page 25: Air2Care In & Hybrimatt 100 In Compressor (Scu)

    7.3 Air2Care IN & HybriMatt 100 IN Compressor (SCU) The Air2Care IN & HybriMatt 100 IN compressor is compatible with Air2Care IN mattresses and with HybriMatt 100 IN mattress. Fig. SCU top side with 3 Air Connectors 1. Three Air Connectors for Three Air Hoses 2.
  • Page 26: Fig. Scu Side View (From The Head End Of The Bed Towards The Foot End When The Scu Is Integrated To The Bed)

    Fig. SCU side view (from the head end of the bed towards the foot end when the SCU is integrated to the bed) Fig. SCU side view (from the foot end of the bed towards the head end when the SCU is integrated to the bed) D9U00A2CI-0101_02...
  • Page 27: Installation Of The 3 Air Hoses To The Air2Care In & Hybrimatt 100 In Compressor

    7.3.1 Installation of the 3 Air Hoses to the Air2Care IN & HybriMatt 100 IN com- pressor Fig. Installation of the 3 Air Hoses to the Air2Care IN & HybriMatt 100 IN compressor Connect 3 Air Pipes to SCU when observing colour code.
  • Page 28: Fixation Of The Air2Care In Mattresses On The Essenza 300 Mattress Support Platform

    7.3.2 Fixation of the Air2Care IN mattresses on the Essenza 300 mattress supp- ort platform Air2Care IN mattresses are designed for Essenza 300 mattress support platform. Air2Care IN mattresses are equipped with straps intended for fixing mattress on the Essenza 300 mattress support platform. LEFT SIDE RIGHT SIDE Fig.
  • Page 29: Pressure Measurement

    7.3.4 Pressure Measurement WARNING! The pressure measurement table below should only be taken as a reference when selecting the optimum wei- ght pressure level. The auto calculated pressure value may be different to the reference setting in the pressure measurement  table. This is due to variation in individual patient’s BMI and posture. Provided the patients weight does not exceed the safe working load of the mattress type that is being used, also the patients BMI is in the normal range for their age and gender, the mattress will prevent the patient from bottoming out even if the weight setting does not correspond to the values in the pressure measurement table.
  • Page 30: Function Description

    8 FUNCTION DESCRIPTION Air2Care IN & HybriMatt 100 IN Control Panel Fig. Keyboard of the Air2Care IN & HybriMatt 100 IN Control Panel 1. APT Mode LED Indicator 2. CLP Mode LED Indicator 3. MAX Mode LED Indicator 4. Pressure Level Indicators 5.
  • Page 31: Auto Pressure Setting (8 - For Air2Care In Mattresses Only)

    8.1 AUTO Pressure Setting (8 - for Air2Care IN mattresses only) WARNING! Visual and hand checks should always be performed by a suitably trained clinical professional if there is any concern that the patient is or may bottoming out! CAUTION! When a patient’s condition has significantly changed, the system will reassess the comfort weight setting ...
  • Page 32: Apt Mode (1)

    8.3 APT Mode (1) Press the MODE Button to enter APT Mode (1 : 2 alternating pressure cell cycle). The Green LED light will indicate the selection of the mode. Caregivers may adjust mattress comfort level by pressing — (6) OR + (7) button. 8.4 CLP Mode (2) Press MODE Button to enter CLP constant low pressure therapy mode.
  • Page 33: Power Failure Alert (12)

    8.10 Power Failure Alert (12) Should there be power failure outage, the Mains Power Fault LED Indicator (12) will light up and the Power Failure alert will activate immediately. Press the MUTE Button on the Control Panel to deactivate the audible alert. When power is back, the ON/OFF Button (10) must be pressed for two seconds to turn on the compressor.
  • Page 34: Installation

    9 INSTALLATION CAUTION! Incorrect use due to incorrect installation or incorrect removal. ► Do not change order of the installation steps and removal steps! Remove any mattress from the mattress support platform of the Essenza 300 medical bed or the Essenza 300 with Lateral Tilt medical bed.
  • Page 35: Fig. Orientation Of The Scu In The Calfrest Frame (Essenza 300 Medical Bed With Lateral Tilt)

    Fig. Orientation of the SCU in the calfrest frame (Essenza 300 medical bed with Lateral Tilt) D9U00A2CI-0101_02...
  • Page 36: Fig. Access To The Power Switch (Essenza 300 Medical Bed In Bottom View)

    When standing opposite the calfrest at foot end of the bed, turn on the Power Switch on the bottom left side of the integrated compressor (SCU). For better access to the Power Switch, set the bed to the highest position. It should be done once only during the first installation.
  • Page 37: Fig. Access To The Power Switch (Essenza 300 Medical Bed With Lateral Tilt In Bottom View)

    Fig. Access to the Power Switch (Essenza 300 medical bed with Lateral Tilt in bottom view) Ensure the selected mattress is compatible with the Control Panels in siderails of the Essenza 300 bed. Put Air- 2Care 6 IN mattress or Air2Care 8 IN mattress or HybriMatt 100 IN mattress on the mattress support platform of the bed.5 Connect Fowler Boost cable from the mattress to the integrated compressor (SCU) under the mattress.
  • Page 38: Fig. Position Of The Scu Under The Calfrest Plastic Cover (Essenza 300 Medical Bed)

    Fig. Position of the SCU under the Calfrest Plastic Cover (Essenza 300 medical bed) Fig. Position of the SCU under the Calfrest Plastic Cover (Essenza 300 medical bed with Lateral Tilt) D9U00A2CI-0101_02...
  • Page 39: Removal

    10 REMOVAL 1)  Turn off the compressor via the ON/OFF Button on the control panel. Ensure the mattress support platform is in horizontal position. Disconnect the 3 Air Hoses from the integrated compressor (SCU). Disconnect Fowler Boost cable from the integrated compressor (SCU). Remove Air2Care 6 IN mattress or Air2Care 8 IN mattress or HybriMatt 100 IN mattress from the mattress support platform of the bed.
  • Page 40: Operating Instructions

    ►   Prior to more than 30 minutes of the Passive Transport Mode, consult qualified hospital staff. The Passive Transport Mode is intended for transporting a patient in bed on the mattress. When Air2Care IN mattress is disconnected from power supply, mattress valve will get locked to keep the mattress inflated during transport. Air2Care  IN mattresses will hold pressure for 30 minutes. Before disconnecting the mattress, it is recommended to adjust mattress MAX pressure mode to reduce risk of patient's bottoming out.
  • Page 41: Passive Transport Mode Of The Hybrimatt 100 In Mattress

    To change the HybriMatt 100 IN mattress to Transport Mode: ► Unlock the Air2Care IN & HybriMatt 100 IN Control Panel on the Essenza 300 foot plastic siderail by Lock Button. ► Turn off the integrated compressor by the ON/OFF button on the Air2Care IN & HybriMatt 100 IN Control Panel. Unplug the Essenza 300 bed from the mains.
  • Page 42: Supine Nursing With Air2Care 6 In And Air2Care 8 In Mattresses

    11.4 Supine Nursing with Air2Care 6 IN and Air2Care 8 IN mattresses WARNING! Risk of injury due to incorrect use! ►   Ensure adequate patient support is maintained during normal mattress operation! Remember that  APT Mode alternately deflates cells in a 1 – 2 cell cycle, hence it is recommended not to manually deflate more  than one cell in mattress head zone and in mattress foot zone! Two adjacent deflated cells lead to the risk of  pressure injuries development! It is forbidden to deflate any mattress cell in the mattress trunk zone! 11.5 Prone Nursing with Air2Care 6 IN and Air2Care 8 IN mattresses WARNING! Risk of injury due to incorrect use! ►  ...
  • Page 43: Correct Use Of The Air2Care In Mattresses With Air2Care In & Hybrimatt 100 In Compressor

    11.8 Correct Use of the Air2Care IN mattresses with Air2Care IN & Hybri- Matt 100 IN compressor Intension Intended User Procedure Installation of Air2Care 6 IN or Air- Caregiver 1. Secure correct mattress position. 2Care 8 IN mattress on the mattress 2.
  • Page 44 2Care 6 IN or the Air2Care 8 IN mattress is connected to the integrated compressor. Pressure in the whole mattress increases about 5 mm Hg during this effect. 3. Fowler Boost LED Indicator on the Air2Care IN & Hyb- riMatt 100 IN Control Panel is lit. CPR of Air2Care 6 IN or Air2Care 8 Caregiver 1.
  • Page 45 Intension Intended User Procedure Cleaning of Air2Care 6 IN or Air2Ca- Caregiver, technical 1. Unlock the Air2Care IN & HybriMatt 100 IN Control re 8 IN mattress personnel Panel on the Essenza 300 foot plastic siderail by Lock Button. 2. Turn off the integrated compressor by the ON/OFF ...
  • Page 46: Correct Use Of The Hybrimatt 100 In Mattress With Or Without Air2Care In & Hybrimatt 100 In Compressor

    2. Turn on the integrated compressor under the bed Calfrest by Power Switch. 3. Press and hold ON/OFF button on the Air2Care IN & HybriMatt 100 IN Control Panel for 2 seconds. An acous- tic signal will beep.
  • Page 47 2. Ensure the Fowler boost cable is connected into the mattress integrated compressor. 3. Ensure the CPR valves are closed. 4. Press and hold ON/OFF button on the Air2Care IN & HybriMatt 100 IN Control Panel for 2 seconds. An acous- tic signal will beep. 5. Mattress starts to inflate. Mode LED Indicators and ...
  • Page 48 Intension Intended User Procedure Cleaning of the HybriMatt 100 IN Caregiver, technical 1. Unlock the Air2Care IN & HybriMatt 100 IN Control mattress personnel Panel on the Essenza 300 foot plastic siderail by Lock Button. 2. Turn off the integrated compressor by the ON/OFF  button on the Air2Care IN & HybriMatt 100 IN Control Panel.
  • Page 49 Intension Intended User Procedure Mattress Cover Replacement Technical personnel 1. Unlock the Air2Care IN & HybriMatt 100 IN Control who is trained to servi- Panel on the Essenza 300 foot plastic siderail by Lock ce HybriMatt 100 IN Button. mattress 2. Turn off the integrated compressor by the ON/OFF ...
  • Page 50: Cleaning Guidelines

    12 CLEANING GUIDELINES Follow the below procedure to clean and decontaminate the system. It is important to follow these steps before using the system again. Cleaning task is required at least once a week to maintain personal hygiene. 12.1 System Control Unit (SCU) ►...
  • Page 51: Maintenance

    13 MAINTENANCE WARNING! Risk of injury when working on the mattress system! ► Ensure that the mattress system is disconnected from the mains power prior to installation, putting into servi- ce, maintenance and deinstallation. WARNING! Risk of injury due to defective mattress system! ►...
  • Page 52: Electrical Safety Check And Preventive Maintenance

    13.3 Electrical Safety Check and Preventive Maintenance 13.3.1 Electrical Safety Check WARNING! Risk of injury due to incorrect electrical safety checks! ► Ensure that electrical safety checks are performed exclusively by manufacturer´s customer service or by authorised service personnel certified by the manufacturer. ►...
  • Page 53: Fuse Change

    Examine if there is any air leakage between the SCU and the mattress connections or from the air mattress tubes: ► Check connectors between the air mattress and the SCU. If there is any disconnection, please reconnect it. ► Check the air-connecting tubes. Ensure each single cell is not broken. ►...
  • Page 54: Troubleshooting

    14 TROUBLESHOOTING Problems Reasons Maintenance 1. After powered on, check if visible LED light turns on. If not, please check the 1. SCU does not work. below issues: 1.1 Check if power cord is plugged into appropriate voltage AC outlet. 1.2 If problem is still present, contact service department approved by manu- facturer!
  • Page 55: Symbols

    15 SYMBOLS Symbol Meaning Power ON. Power OFF. Alternating current. Manufacturer. Manufacturing Date. Serial Number. Type B Applied Part. Read instructions for use. Degree of protection against harmful ingress of water and particulate matter. IP44 D9U00A2CI-0101_02...
  • Page 56 Symbol Meaning CE MARK (EUROPEAN CONFORMITY). GENERAL WARNING SIGN. CAUTION SIGN. DO NOT IRON. TUMBLE DRY ON LOW HEAT SETTING (MAX. 60°C). MACHINE WASH AT MAX. 95°C FOR 3 MINUTES. DISPOSAL OF ELECTRICAL & ELECTRONIC EQUIPMENT (WEEE): THIS PRODUCT SHOULD BE HANDED OVER TO AN APPLICABLE COLLECTION POINT FOR THE RECYCLING OF ELECTRICAL AND ELECTRONIC EQUIPMENT.
  • Page 57 Symbol Meaning DO NOT WRING. DRY. COVER MATERIALS ARE FIRE RESISTANT TO BS7175, SOURCE 0, 1 AND 5. HANDWASH WITH DETERGENT (INITIAL TEMPERATURE OF HOT WATER SHOULD NOT EXCEED 50°C). DISINFECT USING SOLUTION CONTAINING LESS THAN 1000 PPM OF CHLORINE (REFER TO INSTRUCTIONS FOR USE).
  • Page 58: Acoustic Alerts And Signals

    Symbol Meaning ON/OFF BUTTON (PRESS THE ON/OFF BUTTON TO TURN ON/OFF THE COMPRES- SOR.) DISINFECT USING SOLUTION CONTAINING LESS THAN 10000 ppm OF CHLORINE (REFER TO INSTRUCTIONS FOR USE). MACHINE WASH AT MAX. 71°C FOR 3 MINUTES. 16 ACOUSTIC ALERTS AND SIGNALS Single beep sounds after pressing a button when corresponding function is not supported with current mattress or function is not supported on current setting on the control panel.
  • Page 59: Technical Specifications

    17 TECHNICAL SPECIFICATIONS All technical data are rated data and are subject to construction and manufacturing tolerances. Air2Care IN & HybriMatt 100 IN integrated mattress system is suitable for continuous operation. 17.1 Mechanical Specifications of the Air2Care IN Mattresses Parameter Value Dimensions Air2Care 6 IN mattress (inflated) - length x width x  200 cm × 86 cm × 15 cm „...
  • Page 60: Mechanical Specifications Of The Hybrimatt 100 In Mattress

    Value External Dimensions of System Control Unit (length x width 22 x 39,5 x 13,5 cm x height) Weight of the Air2Care IN & HybriMatt 100 IN System Control 2 kg Unit Sound Pressure Level of System Control Unit 35 dBA...
  • Page 61: Electrical Specifications Of The Air2Care In & Hybrimatt 100 In Compressor

    WARNING! Risk of damaging the product due to incorrect environment conditions! ► Do not use the Air2Care IN Integrated Mattress System under the environmental conditions outside of those specified in the Air2Care IN Environment Conditions chapter! CAUTION! Risk of damaging the product if its packaging is exposed to environmental conditions outside of those speci-...
  • Page 62: Hybrimatt 100 In Environment Conditions

    17.6 HybriMatt 100 IN Environment Conditions WARNING! Risk of damaging the product due to incorrect environment conditions! Do not use the HybriMatt 100 IN Integrated Mattress System under the environmental conditions out- ► side of those specified in the HybriMatt 100 IN Environment Conditions chapter! CAUTION! Risk of damaging the product if its packaging is exposed to environmental conditions outside of those speci- fied in the HybriMatt 100 IN Environment Conditions chapter!
  • Page 63: Electromagnetic Compatibility

    WARNING! Mobile RF communication device (including end use devices like antenna cables and external antenna) should not be used closer than 30 cm (12 inches) from any part of this Air2Care IN & HybriMatt 100 IN integrated mattress system, including cables specified by manufacturer. Otherwise this could lead to deterioration of  functionality of this Air2Care IN & HybriMatt 100 IN integrated mattress system.
  • Page 64: Manufacturer Instructions - Electromagnetic Emissions

    18.2 Manufacturer Instructions - Electromagnetic Emissions Immunity Tests Compliance level Electrostatic discharge (ESD) ± 8 kV for contact discharge IEC 61000-4-2 ± 15 kV for air discharge Radiated RF 3 V/m IEC 61000-4-3 80 MHz – 2,7 GHz 80 % AM at 1 kHz Proximity fields from RF wireless communications equipment IEC 61000-4-3 See Table 1...
  • Page 65: Disposal

    Based on the Directive No. 2002/96/ EC (Directive WEEE - Waste, Electric and Electronic Equipments) the company LINET, s. r. o. is registered in the List of Electric and Electronic Equipment Producers (Seznam výrobců elektrozařízení) on the Ministry of the Environment of the Czech Republic...
  • Page 66: Warranty

    WARNING! Follow the national requirement to dispose unit! LINET ® will only be held responsible for the safety and reliability of products that are regularly serviced, maintained and used in accordance with the safety guidelines. Should a serious defect arise that cannot be repaired during maintenance: ►...