Troubleshooting By Guide - VWR Mega Star 600 Instruction Manual

Hide thumbs Also See for Mega Star 600:
Table of Contents

Advertisement

Troubleshooting by Guide

If problems occur other than those listed in this table, the authorized customer service representative must
be contacted.
Error number
Error message
E-002;
Read Manual
E-005;
E-008;
E-010;
E-011;
E-012;
E-015;
E-016;
E-034;
E-036;
E-041;
E-048;
E-050;
E-051;
E-052;
E-053;
E-054;
E-072;
E-077;
E-101;
E-104
E-031
Temp High!
E-017;
Read Manual
E-020;
E-021;
E-022;
E-023
E-078;
E-079;
E-080;
E-081;
31
Troubleshooting
Restart the centrifuge.
If the error message appears again, inform the service.
CAUTION
Hot metal parts!
Check, if the centrifuge is accessible.
Be sure, that the room temperature is within the limits.
Let the centrifuge cool down for 15 minutes.
Be sure there is no condensed water in the rotor chamber.
If the error message appears again, inform the service.
Wait until the rotor has stopped.
Check, if the rotor is qualified for the VWR Mega Star 600 / 600R
Centrifuge.
Check, if the bottom of the rotor is damaged and if the rotor is placed
on the Auto-Lock correctly.
If the error message appears again, inform the service.
VWR Mega Star 600 / 600R ver b, rel 17.07.2020

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vwr mega star 600rVwr mega star 600r micro521-1893521-1894521-1895