Aerosol-tight rotors

For your protection when using an aerosol tight lid, the rotor should only be removed with the lid closed.
NOTICE Rotors supplied with a lid for aerosol-tight applications come with a mandrel, which belongs to the
Auto-Lock. Be sure not to place the lid onto this mandrel to prevent it from being damaged.
CAUTION Skin can be pierced by sharp mandrel tip.
Do not touch the mandrel.

Switch off Centrifuge

To turn off the centrifuge push the mains switch to "0".

System Menu

To get into the system menu, press and hold any key on the front panel turning on the centrifuge. Keep this
key depressed until system menu appears in display. Navigate through the system menu using the TIME
arrow keys.

Flowchart System Menu

The system menu can be navigated using the SPEED arrow keys. The shown entry can be changed using
the TIME and TEMP arrow keys. Press the START key to save this edit and quit the system menu.
Press the STOP key to quit the system menu.
Values shown at some entries in the picture below are only examples.
NOTICE A total number of cycles completed on the centrifuge is counted, since installation or a new main
board was installed.
Cycle counts should be periodically recorded to help with the identification of the total number of runs a rotor
has completed.