Service manual M500 N, M520, M525
Doc Code 10 665 751
CAN-handles: show parameters
DI-controller: show parameters
DI-controller: test shutters
DI-LCD-module: show parameters
IGS-Workstation: show parameters
Test and diagnosis in the user menu:
Select [Menue II]
Select [User settings]
Available test function
CAN-Handle left/right
Foot/handswitch connector 1 / 2
Microscope
Control unit
Stand
Restore standard settings*
*not visible when standard settings
are activated
4-13
HW and SW version of left and right CAN handle are displayed.
Status and version of a connected DI C500 module are displayed.
Shutters of a connected DI C500 module can be activated and monitored.
Status and version of the LCD module of a connected DI C500 module are
displayed.
Status and version of a connected IGS system are displayed.
Description
All keys of the appropriate handle can be checked; the standard
setting of the handles can be displayed.
All keys of the appropriate foot/handswitch can be checked; the standard
setting of the foot/handswitch can be displayed.
Speed and position of focus or zoom can be displayed.
Setting of screen brightness and contrast ca be displayed.
All keys of the stand handgrips (Not CAN-handles!) can be checked; the
standard configuration of the handgrip can be displayed.
Overwrite the actual user with default settings.
Fault finding
08/2007 Gahe