Leica M500 N Service Manual page 35

Table of Contents

Advertisement

Service manual Leica M500 N, M520, M525
Doc Code 10 665 751
Stock n°.
Description / Specification
Service programs – Leica M500N, M520, M525
Service programs in control unit
How to enter the service functions:
Enter the service menu of the M500 N/M520 control unit in menue II
Enter security code: 6642
Select the service functions
Validity
Below description refers to SW 1.87 in the control unit and 2.4 or 4.4 in optics carrier.
Service function
Standard and user configuration
Ad. Function 1 output configuration
Ad. Function 2 output configuration
Screen saver setting
User parameter save time setting
Power off timer setting
3-1
Remark
Description
Settings of user- and standard
configuration can be changed and
stored.
The output of additional function 1 can
be set and stored either to toggle or to
switch mode.
The output of additional function 2 can
be set and stored either to toggle or to
switch mode.
Screen saver for control unit display:
after the selected time the brightness
will be reduced.
When user settings are saved, the
display shows "User setting saved at..."
for a certain time; this time can be
altered.
A power off / power on within the
selected time will effect:
The last selected user (before switching
off) is automatically taken as the actual
user.
Service programs
08/2007 GAHE
Remark
Refer to
stand user
manual.
Switch mode: relais contact closes
as long as the release button is
pressed.
Toggle mode: relais contact
opens/closes with each activation of
the release button.
-"-
Adjustable between 0..60 minutes.
Default = 5 minutes.
0 minutes will disable screen saver.
Adjustable between 2..6 seconds.
Default = 4 seconds.
Adjustable between 0..60 minutes.
Default = 5 minutes.
Drawing