Service manual Leica M500 N, M520, M525
Doc Code 10 665 751
Stock n°.
Description / Specification
Service programs – Leica M500N, M520, M525
Service programs in control unit
How to enter the service functions:
•
Enter the service menu of the M500 N/M520 control unit in menue II
•
Enter security code: 6642
•
Select the service functions
Validity
Below description refers to SW 1.87 in the control unit and 2.4 or 4.4 in optics carrier.
Service function
Standard and user configuration
Ad. Function 1 output configuration
Ad. Function 2 output configuration
Screen saver setting
User parameter save time setting
Power off timer setting
3-1
Remark
Description
Settings of user- and standard
configuration can be changed and
stored.
The output of additional function 1 can
be set and stored either to toggle or to
switch mode.
The output of additional function 2 can
be set and stored either to toggle or to
switch mode.
Screen saver for control unit display:
after the selected time the brightness
will be reduced.
When user settings are saved, the
display shows "User setting saved at..."
for a certain time; this time can be
altered.
A power off / power on within the
selected time will effect:
The last selected user (before switching
off) is automatically taken as the actual
user.
Service programs
08/2007 GAHE
Remark
Refer to
stand user
manual.
Switch mode: relais contact closes
as long as the release button is
pressed.
Toggle mode: relais contact
opens/closes with each activation of
the release button.
-"-
Adjustable between 0..60 minutes.
Default = 5 minutes.
0 minutes will disable screen saver.
Adjustable between 2..6 seconds.
Default = 4 seconds.
Adjustable between 0..60 minutes.
Default = 5 minutes.