Service manual M500 N, M520, M525
Doc Code 10 665 751
Error message
Timeout of autofocus function
Set CAN Handle R in device list
Set CAN Handle L in device list
CAN Handle R not detected
CAN Handle L not detected
Set DI-Controller in device list
DI-Controller not detected
DI-Controller shutter position error
DIC500: Video signal not valid
DIC500: detected (config. error)
DIC500: not detected (config. error)
Imaging position sensor error
Stand not available
Stand is busy
Stand error pending, enter diagnose Setup menu of stand was called up,
4-11
Description
Autofocus function was started, but no
response about the cycle termination
within 36 second.
CAN Handle R was found on CAN bus,
although it is set to NOT PRESENT.
CAN handle L was found on CAN bus,
although it is set to NOT PRESENT.
CAN handle R was not found on CAN
bus, although it is set to PRESENT.
CAN handle L was not found on CAN
bus, although it is set to PRESENT.
DI-Controller was found on CAN bus,
although it is set to NOT PRESENT.
DI-Controller was not found on CAN bus,
although it is set to PRESENT.
Detected position of shutter does not
correlate with commanded position for
more than 5 seconds.
Video signal not valid or PLL not
engaged.
DI-Color LCD module was found on CAN
bus, although DI controller is set to NOT
PRESENT.
DI-Color LCD module was not found on
CAN bus, although DI controller is set
PRESENT.
Image position sensor indicates that
prism is not in left or right end position.
Setup- or diagnostic menu of stand was
called up, but stand does not respond.
Setup- or diagnostic menu of stand was
called up while stand is busy (for
example during auto balance
procedure).
although an error of the stand is
pending.
Fault finding
08/2007 Gahe
Remedy
Problem in autofocus devices
Set CAN handle R in device list
PRESENT.
Set CAN handle L in device list
PRESENT.
Set CAN handle R in device list
NOT PRESENT.
Set CAN handle L in device list
NOT PRESENT.
Set DI-Controller in device list
PRESENT.
Set DI-Controller in device list
NOT PRESENT.
Problem in DI-Controller / DI C
module.
Check video signal
Check VESA settings
Set DI controller in device list
PRESENT.
Set DI controller in device list NOT
PRESENT.
Engage rotary knob for
image left/ right to left or
right end position.
Else defective sensor in DI-
C module.
Stand not present
Failure in CAN cabling
Failure in stand CAN
controller
Wait until stand is ready and repeat
call up.
Confirm stand error in diagnostic
menu and repeat call up.