Service programs
08/2007 GAHE
Stock n°
Description / Specification
Service function
User selection timer setting
Tube focal length setting
Exception reboot timer setting
Set device list
Restore factory defaults
Autofocus setup
Stand setup
DI Default Settings
DI Color LCD Module Settings
3-2
Remark
Description
During the power up routine the control
unit allows to select a user; if no user is
selected within a certain time it will take
the standard settings.
The setting influences the displayed
magnification value in the main menu.
If an error causes an exception reboot of
the control unit, the error can be
displayed for a certain time before the
reboot starts.
Connected CAN devices have to be set
to "present".
Not connected CAN devices have to be
set to "Not present".
All standard and user settings will be
reset to the factory default values.
Autofocus video parameters and
autofocus range parameters are set in
this menu.
Opens a monitor for entering setup /
service program of the connected stand
system.
Shutter functions and shutter mode are
set in this menu.
Video parameter for the LCD display of
the DI-module are set in this menu.
Service manual Leica M500 N, M520, M525
Doc Code 10 665 751
Remark
Adjustable between 2..20 seconds.
Default = 11 seconds.
Adjustable between 100..200.
Set to focal length of main tube.
Default = 170.
Adjustable between 0..20 seconds.
Default = 10 seconds.
If the device list is not set correctly, it
may result in malfunction and/or
error messages. Available CAN
devices:
Stand (M500-N)
Microscope (M500-N)
Microscope (M500-O)
X/Y-Unit (M500-O)
Autofocus (M500-N)
Handle R (M500-N)
Handle L (M500-N)
DI control (M500-N)
Swing carrier (MS3)
All user settings will be lost.
Details in
service manual
BIV/Autofocus 10 665
Details in the appropriate stand
service manual.
Details in
service manual Neuro
modules 10665683.
Details in
service manual Neuro
modules 10665683.
Drawing
645.